4250 Flow Meter
Section 2 Programming
2-33
chloride) and sea water. If you use the value of 1.91, in most cases
you will be able to identify that gross changes are occurring in
the ionic content of the stream. If you move with the
right
arrow
, the following options will appear on the display:
T.D.S. stands for “total dissolved solids.” T.D.S. are measured in
parts per thousand (ppt).
Total dissolved solids are estimated by multiplying conductivity
by an empirical factor. This factor can vary between 0.55 and 0.9
depending on the solubility of the ions in the water and its tem-
perature.
Select the appropriate temperature units.
2.4.4 Step 2 - Flow
Conversion (Area
Velocity Only)
Flow conversion is the method the flow meter uses to calculate
flow rate from the measurement of the area and velocity of the
flow stream.
Note
Accurate measurements of both area and level are essential
for accurate flow compilation. See
Section
, Area Velocity
Sensor Installation.
The 4250 can calculate flow either through area-velocity mea-
surements and area-velocity data points, or through level-to-flow
rate conversions using a standard primary measuring device or
level-to-flow rate data points.
If you wish to use the 4250 with a primary measuring device,
skip to
Step 2 - Flow Conversion (Level-to-Flow Rate)
fol-
lowing this section.
If you select Area Velocity the following will appear:
CHANNEL SHAPE
uses the shape of the channel to calculate
the flow rate. DATA POINTS allows you to enter your own mea-
surements of area and velocity and the flow meter will calculate
flow rate based on your data. For now, select CHANNEL SHAPE.
YSI 600 CONDUCTIVITY PARAMETER
• YSI T.D.S. • • NOT MEASURED •
TDS SCALE FACTOR
0.75
YSI 600 TEMPERATURE UNITS
• °F • •°C • • NOT MEASURED •
FLOW CALCULATION
• AREA VELOCITY • • LEVEL TO FLOW RATE •
AREA VELOCITY CALCULATION
• CHANNEL SHAPE• • DATA POINTS •
Содержание 4250
Страница 2: ......
Страница 4: ......
Страница 12: ...4250 Flow Meter Table of Contents x ...
Страница 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Страница 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Страница 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Страница 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Страница 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Страница 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Страница 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Страница 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Страница 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ...Warranty ...
Страница 204: ......