4250 Flow Meter
Section 4 Options and Accessories
4-7
Table 4-2 Multiple Analog Output Board Specifications
Operating Temperature
0 to 140° F (–18 to 60° C)
Output Accuracy
0.5% of full-scale
Resolution 0.1%
of
full-scale (0-20 mA)
Electrical Isolation
Isolated from each other and from the flow meter.
Calibration
Factory-calibrated; no further adjustments necessary.
Maximum output distance
1500 feet (457.3 m) using 18 AWG wire.
Current Range (per loop)
0 to 20 mA
Maximum Load (per loop)
750 ohms
Analog Output 1 Terminations
Pin A
(Red wire –)
Pin C
(White wire +)
Analog Output 2 Terminations
Pin D
(Brown wire –)
Pin F
(Blue wire +)
Analog Output 3 Terminations
Pin E
(Black wire –)
Pin B
(Green wire +)
Note: If you must strip the cable further back to facilitate wiring, you will expose an orange, yellow,
and purple wire. You may disregard these wires, as they are not connected in this application.
Содержание 4250
Страница 2: ......
Страница 4: ......
Страница 12: ...4250 Flow Meter Table of Contents x ...
Страница 28: ...4250 Flow Meter Section 1 Introduction 1 16 ...
Страница 90: ...4250 Flow Meter Section 2 Programming 2 62 ...
Страница 140: ...4250 Flow Meter Section 4 Options and Accessories 4 30 ...
Страница 160: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 2 ...
Страница 161: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 3 ...
Страница 166: ...4250 Flow Meter Appendix A Replacement Parts and Accessories A 8 ...
Страница 184: ...4250 Flow Meter Appendix C General Safety Procedures C 10 ...
Страница 191: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 7 ...
Страница 192: ...4250 Flow Meter Appendix D Material Safety Data Sheets D 8 ...
Страница 196: ......
Страница 198: ......
Страница 200: ......
Страница 202: ...Warranty ...
Страница 204: ......