Teledyne API Model 401 O
3
Photometric Calibrator Instruction Manual, 01124, Rev. J1
1.4 Specifications
Photometer
Ranges
User selectable to any full scale range from 100 ppb to 20,000 ppb
Zero Noise
< 0.3 ppb (rms)
Span Noise
< 0.5% of reading (rms)
Lower Detectable Limit
< 0.6 ppb (rms)
Zero Drift (24 hours)
< 1.0 ppb*
Zero Drift (7 days)
< 1.0 ppb*
Span Drift (24 hours)
1% of reading *
Span Drift (7 days)
2% of reading *
Linearity
Better than 1% FS
Precision
1.0 ppb
Lag Time
10 sec
Rise/Fall Time (95%)
<20 sec
Sample Flow Rate
800 cc/min
±
10%
Temperature Range
5-40
°
C
Humidity Range
0-95% RH, Non-Condensing
Temp Coefficient
< 0.05 % per
°
C
Voltage Coefficient
< 0.05 % per V
Dimensions HxWxD
7" x 17" x 27" (178 mm x 432 mm x 686 mm)
Weight, Analyzer
60 lb. (27 kg)
Power, Analyzer
110v/60Hz, 220v/50Hz 250 watts
Environmental Conditions
Installation Category (Overvoltage Category) II
Pollution Degree 2
Recorder Instantaneous
±
100 mV,
±
1 v
(Bi-Polar)
±
5 v,
±
10 v
Status
12 Status Outputs from Opto-Isolators
Calibration Ozone Generator
Flow rate
2 to 5 LPM adjustable
Maximum concentration
1.0 ppm @ 5 LPM
Minimum concentration
0.050 ppm @ 2 LPM
Stability
±
0.002 ppm (bench feedback)
* at constant temperature and voltage
1-5
Содержание 401
Страница 40: ...Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual 01124 Rev J1 INTENTIONALLY BLANK 2 10 ...
Страница 46: ...Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual 01124 Rev J1 INTENTIONALLY BLANK 3 6 ...
Страница 54: ...Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual 01124 Rev J1 INTENTIONALLY BLANK 4 8 ...
Страница 58: ...Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual 01124 Rev J1 INTENTIONALLY BLANK 6 2 ...
Страница 100: ...Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual 01124 Rev J1 INTENTIONALLY BLANK 10 4 ...
Страница 106: ...Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual 01124 Rev J1 INTENTIONALLY BLANK A 2 ...
Страница 108: ...Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual 01124 Rev J1 INTENTIONALLY BLANK B 2 ...
Страница 110: ...Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual 01124 Rev J1 INTENTIONALLY BLANK C 2 ...
Страница 114: ...Teledyne API Model 401 O3 Photometric Calibrator Instruction Manual 01124 Rev J1 INTENTIONALLY BLANK E 2 ...