340CF ELSD
Section 3 Installation
3-9
3.4.1 Operating
Temperatures
3.4.2 Shut Down Procedure
For HPLC applications, the ELSD must count down to the
standby state.
Press
MENU
|
POWER
until the standby countdown begins. The
unit will enter standby at the end of five minutes. Once the unit
is in the standby state, turn off the laser and heaters.
If there are no leaks between your gas source and the ELSD, you
do not have to turn the gas off at the source.
Table 3-1 Suggested Operating Temperatures for HPLC Gradient Separations
Solvent @ 1.0mL/min.
Drift Tube Temperature, °C
Spray Chamber Temperature, °C
90/10 Water/METHANOL
45
15
90/10 Methanol/WATER
65
50
50/50 Methanol/Water
60
30
Acetonitrile
60
50
Methanol
60
50
Water
45
10