Operating Instructions
Model 200E Instruction Manual
The second operating mode is the
SETUP
mode. This mode is used for configuring various
features and functions of the analyzer, such as the iDAS system, the analog output ranges,
or the COM port settings. The
SETUP
mode is also used for performing various diagnostic
tests during troubleshooting.
The third operating mode is the
CAL
mode, which allows calibration of the analyzer in
various ways. Because of its importance, this mode is described separately in Chapter 7.
The mode field (upper left corner) of the front panel display indicates the current operating
mode (refer to Figure 3-4 for front panel features). In addition to the three main modes,
there are mode variations, which are summarized in Table 6-1.
6.2. Sample Mode
This is the analyzer’s standard operating mode. In this mode, the instrument is analyzing
NO and NO
X
and calculating NO
2
concentrations.
6.2.1. Test Functions
A series of test functions is available at the front panel while the analyzer is in
SAMPLE
mode. These parameters provide information about the present operating status of the
instrument and are useful during troubleshooting (Section 11.1). They can also be recorded
in one of the iDAS channels (Section 6.10.1.1) for data analysis. To view the test functions,
press one of the
<TST TST>
keys repeatedly in either direction.
NOTE
A value of “XXXX” displayed for any of the TEST functions indicates an out-of-
range reading or the analyzer’s inability to calculate it.
NOTE
All pressure measurements are represented in terms of absolute pressure.
Absolute, atmospheric pressure is 29.92 in-Hg-A at sea level. It decreases about 1
in-Hg per 300 m gain in altitude. A variety of factors such as air conditioning and
passing storms can cause changes in the absolute atmospheric pressure.
42
044100102 Rev A
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...