Model 200E Instruction Manual
Getting Started
3.5.4. Functional Check
After the analyzer’s components have warmed up for at least 30 minutes, verify that the
software properly supports any hardware options that were installed.
Check to make sure that the analyzer is functioning within allowable operating parameters.
Appendix A and C include a list of test functions viewable from the analyzer’s front panel as
well as their expected values. These functions are also useful tools for diagnosing perform-
ance problems with your analyzer (Chapter 11). The enclosed
Final Test and Validation Data
Sheet
(part number 04490) lists these values before the instrument left the factory. To
view the current values of these test functions press the <TST TST> keys:
RANGE
NOX STB
SAMP FLW
OZONE FL
PMT
NORM PMT
AZERO
HVPS
RCELL TEMP
BOX TEMP
PMT TEM
IZS TEMP
1
MOLY TEMP
RCEL
SAMP
NOX SLOPE
NOX OFFS
NO SLOPE
NO OFFS
NO2
2
NOX
2
NO
2
TEST
3
TIME
SAMPLE
RANGE = 500.0 PPB
NOX = X.X
< TST
TST >
CAL
SETUP
1
Only appears if IZS option is
installed.
2
Only appears if analyzer is set
to display NO, NO
2
, NO
x
simultaneously in message
field.
3
Only appears if analog autput
A4 is actively reporting a test
function
Refer to
Section
6.2.1 for
definitions
of these
test
functions.
Toggle
<TST TST>
keys to
scroll through list of functions
Figure 3-11 and Figure 3-12 show the analyzer flow diagrams without and with options
installed. Refer to these diagrams whenever troubleshooting or a thorough understanding of
the analyzer operation is required. Note that valve options for zero/span valves or IZS are
principally the same, except that the span gas stream is connected differently in the
zero/span option.
More detailed pneumatic diagrams for all configurations (base, zero/span valves, IZS,
sample dryer) can be found in Appendix D, diagram number 04574.
044100102 Rev A
23
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...