Getting Started
Model 200E Instruction Manual
The following table includes a brief description of the various warning messages that may
appear after the warm-up time. If warning messages persist after 30 minutes, investigate
their cause using the troubleshooting guidelines in Chapter 11. To view and clear warning
messages, use the following key strokes:
SAMPLE
HVPS WARNING NOX = 0.0
TEST
CAL
MSG
CLR
SETUP
Make sure warning messages are
not due to real problems.
Press
CLR
to clear the current
message.
If more than one warning is active, the
next message will take its place
Once the last warning has been
cleared, the analyzer returns to
SAMPLE
mode
SAMPLE
RANGE=500.0 PPB
NO = 0.0
< TST TST > CAL
MSG
CLR
SETUP
SAMPLE
HVPS WARNING
NOX = 0.0
TEST CAL
MSG
CLR
SETUP
TEST
deactivates warning
messages
MSG
activates warning
messages.
<TST TST
> keys replaced with
TEST
key
NOTE:
If the warning message persists
after several attempts to clear it,
the message may indicate a
real problem and not an artifact
of the warm-up period
Table 3-6: Possible Warning Messages at Start-Up
Message Definition
ANALOG CAL WARNING
The instrument’s A/D circuitry or one of its analog outputs is not
calibrated.
AZERO WRN XXX.X MV
The reading taken during the auto-zero cycle is outside of specified limits.
The value XXX.X indicates the auto-zero reading at the time of the
warning.
BOX TEMP WARNING
The temperature inside the M200E chassis is outside the specified limits.
CANNOT DYN SPAN
Remote span calibration failed while the dynamic span feature was ON
CANNOT DYN ZERO
Remote zero calibration failed while the dynamic zero feature was ON.
CONFIG INITIALIZED
Configuration was reset to factory defaults or was erased.
CONV TEMP WARNING
NO
2
converter temperature is outside of specified limits.
DATA INITIALIZED
iDAS data and settings were erased.
HVPS WARNING
High voltage power supply for the PMT is outside of specified limits.
IZS TEMP WARNING
On units with IZS options installed: The permeation tube temperature is
outside of specified limits.
OZONE FLOW WARNING
Ozone flow is outside of specified limits.
OZONE GEN OFF
Ozone generator is off, which is intentional for the warm-up period. This
is the only warning message that automatically clears itself after warmup.
PMT TEMP WARNING
PMT temperature is outside of specified limits.
RCELL PRESS WARN
Reaction cell pressure is outside of specified limits.
RCELL TEMP WARNING
Reaction cell temperature is outside of specified limits.
REAR BOARD NOT DET
The CPU is unable to communicate with the motherboard.
RELAY BOARD WARN
The firmware is unable to communicate with the relay board.
SAMPLE FLOW WARN
The flow rate of the sample gas is outside the specified limits.
SYSTEM RESET
This message appears every time the analyzer was powered up.
22
044100102 Rev A
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...