
Model 200E Instruction Manual
Getting Started
3.4. Electrical Connections
Refer to Figure 3-3 for the location of the rear panel electrical and pneumatic connections.
3.4.1. Power Connection
Attach the power cord to the analyzer and plug it into a power outlet capable of carrying at
least 10 A current at your AC voltage and that it is equipped with a functioning earth
ground.
The M200E analyzer can be configured for both 100-130 V and 210-240 V at either 50 or 60
Hz., To avoid damage to your analyzer, make sure that the AC power voltage matches the
voltage indicated on the rear panel serial number label and that the frequency is between
47 and 63 Hz. Refer to Figure 3-1 to locate and identify the power configuration plug if
necessary.
If you need to reconfigure your analyzer for one of the other voltage ranges, contact the
factory for an optional power configuration jumper (shown in Figure 3-1)
CAUTION
The correct voltage and frequency needs to be supplied to the analyzer
as listed on the label on the rear panel of the instrument .
Power connections must have a functioning ground connection.
3.4.2. Analog Output Connections
Attach a strip chart recorder and/or data-logger to the appropriate contacts of the analog
output connecter on the rear panel of the analyzer.
ANALOG OUT
A1 A2 A3 A4
+ - + - + - + -
Figure 3-7: Analog Output Connector
Output channels
A1
,
A2
and
A3
are assigned to the
NO
X
, NO and NO
2
concentration
signals of the analyzer. The output labeled A4 can be set by the user for output of any test
parameter accessible through the <TST TST> keys on the front panel display. Section 6.7.4
describes how to change that setting.
Optional current loop outputs are available for analog output channels A1-A3. Pin-outs for
the analog output connector at the rear panel of the instrument are listed in Table 3-2.
The default analog output voltage setting of the M200E is 0 - 5 VDC with a range of
0-500 ppb. To change these settings, see Sections 6.5 and 6.7.3.
044100102 Rev A
17
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...