APPENDIX A
Model 200E Instruction Manual
Table A-3: M200E Test Measurements, Revision C.8
TEST
Function
Message Text
Valid Data Range
Description
RANGE
RANGE=500.0 PPB
50-20000 ppb
Analog output range in single or auto-
range modes.
RANGE1 RANGE1=500.0
PPB
3
50-20000 ppb
D/A #1 range in independent range mode.
RANGE2 RANGE2=500.0
PPB
3
50-20000 ppb
D/A #2 range in independent range mode.
RANGE3 RANGE3=500.0
PPB
3
50-20000 ppb
D/A #3 range in independent range mode.
NOX STB
NOX STB=0.5 PPB
≤
1 ppb with zero air Concentration stability (standard deviation
based on
STABIL_FREQ
and
STABIL_SAMPLES
)
SAMPLE FL
SAMPLE FL= 500
CC/M
500 ± 50 cm³/min Sample flow rate
OZONE FL
Ozone fl= 80 CC/M
80 ± 15 cm³/min FLow rate of ozone generator supply air
PMT
PMT= 000 MV
0-5000 mV for
0-20000 ppb
PMT response
NORM PMT NORM PMT=800.0 MV
0-5000 mV for
0-20000 ppb
Normalized PMT response
AZERO= 40 MV
-20 to 150 V
Auto-zero value, used to normalize PMT
HVPS
HVPS=700 V
400-900 ° C
High voltage power supply voltage of the
PMT.
RCELL TEMP RCELL TEMP= 50.0 C
50 ± 1 ° C
Reaction cell temperature
BOX TEMP
BOX TEMP= 30.0 C
~9 to ~45° C
Box temperature; is about 5° C higher
than ambient temperature
PMT TEMP
PMT TEMP= 7.0 C
7 ± 2 ° C
PMT temperature, cooled with thermo-
electric cooler
MF TEMP
1
MF TEMP=50.8 C
50 ± 1 ° C
Manifold temperature (optional)
IZS TEMP
1
IZS TEMP= 50.0 C
50 ± 1 ° C
Internal zero/span option temperature
MOLY TEMP MOLY TEMP= 315 C
315 ± 5 ° C
NO
2
converter temperature
RCEL
pressure
RCEL= 5.0 IN-HG-A
<10 in-Hg-A
Reaction cell (=vacuum) pressure
SAMP
pressure
SAMP= 29.0 IN-HG-A Ambient - ~1 in-Hg-A Sample pressure, usually about 1 less than
ambient, atmospheric pressure.
NO
X
SLOPE
NOX SLOPE= 0.000
1.0 ± 0.3
Slope of linear fit for NO
X
response,
computed during calibration
NOX OFF
NOX OFF= 0.0 MV
-50 to 150 mV
Offset of linear fit for NO
X
response,
computed during calibration.
NO SLOPE
NO SLOPE= 0.000
1.0 ± 0.3
Slope of linear fit for NO response,
computed during calibration
NO OFFS
NO OFFS= 0.0 MV
-50 to 150 mV
Offset of linear fit for NO response,
computed during calibration.
Time of Day
TIME=15:11:24
00:00:00 to 23:59:59 Current instrument time (24-hour format)
AZERO
242
044100110 Rev A
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...