
Troubleshooting & Repair
Model 200E Instruction Manual
•
Check for disconnected cables to the sensor module.
•
Carry out an electrical test with the
ELECTRICAL TEST
procedure in the diagnostics
menu, see Section 6.7.6. If this test produces a concentration reading, the
analyzer’s electronic signal path is correct.
•
Carry out an optical test using the
OPTIC TEST
procedure in the diagnostics menu,
see Section 6.7.5. If this test results in a concentration signal, then the PMT sensor
and the electronic signal path are operating properly. If the M200E passes both
ETEST and OTEST, the instrument is capable of detecting light and processing the
signal to produce a reading. Therefore, the problem must be in the pneumatics or
the ozone generator.
•
If NO
2
signal is zero while NO signal is correct, check the NO/NO
X
valve and the NO
2
converter for proper operation.
11.3.3. Unstable Zero and Span
Leaks in the M200E or in the external gas supply and vacuum systems are the most
common source of unstable and non-repeatable concentration readings.
•
Check for leaks in the pneumatic systems as described in Section 11.5.1. Consider
pneumatic components in the gas delivery system outside the M200E such as a
change in zero air source (ambient air leaking into zero air line or a worn-out zero
air scrubber) or a change in the span gas concentration due to zero air or ambient
air leaking into the span gas line.
•
Once the instrument passes a leak check, do a flow check (this chapter) to make
sure that the instrument is supplied with adequate sample and ozone air.
•
Confirm the sample pressure, sample temperature, and sample flow readings are
correct and steady.
•
Verify that the sample filter element is clean and does not need to be replaced.
11.3.4. Inability to Span - No SPAN Key
In general, the M200E will not display certain keyboard choices whenever the actual value
of a parameter is outside of the expected range for that parameter. If the calibration menu
does not show a SPAN key when carrying out a span calibration, the actual concentration
must be outside of the range of the expected span gas concentration, which can have
several reasons.
•
Verify that the expected concentration is set properly to the actual span gas
concentration in the
CONC
sub-menu.
•
Confirm that the NO
x
span gas source is accurate. This can be done by comparing
the source with another calibrated analyzer, or by having the NO
x
source verified by
an independent traceable photometer.
•
Check for leaks in the pneumatic systems as described in Section 11.5.2. Leaks can
dilute the span gas and, hence, the concentration that the analyzer measures may
fall short of the expected concentration defined in the
CONC
sub-menu.
•
If the low-level, hardware calibration has drifted (changed PMT response) or was
accidentally altered by the user, a low-level calibration may be necessary to get the
analyzer back into its proper range of expected values. One possible indicator of this
scenario is a slope or offset value that is outside of its allowed range (0.7-1.3 for
200
044100102 Rev A
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...