
Model 200E Instruction Manual
EPA Protocol Calibration
Second, turn on and adjust the O
3
generator in the calibrator to produce sufficient O
3
to
decrease the NO concentration to about 10% of full scale. This will be equivalent to 80% of
the URL of the NO
2
channel. After the analyzer responses stabilize, record the resultant NO,
NO
X
, and NO
2
concentrations.
NO
2
CONVERTER EFFICIENCY
if the NO
X
reading should drop to less than 96% of its starting value during this step, it
indicates the NO
2
converter is in need of troubleshooting or replacement. See Chapter 11
for further details.
While maintaining all other conditions, adjust the ozone generator to obtain several other
concentrations of NO
2
evenly spaced between the 80% URL point and the zero point.
Record the NO, NO
X
, and NO
2
concentrations for each additional point.
Calculate the resulting NO
2
concentrations as follows:
F
]
NO
[
*
F
+
]
[NO
-
]
[NO
=
]
NO
[
T
IMP
2
NO
REM
ORIG
OUT
2
Eq 8-9
Where [NO]
ORIG
is the NO concentration before the GPT ozone is turned on, and [NO]
REM
is
the NO remaining after GPT.
Plot the NO
2
concentration output by the instrument on the y-axis against the generated
NO
2
[NO
2
]
OUT
on the x-axis. The plot should be a straight line within the ± 2% linearity
criteria given for the NO
x
and NO channels. If the plot is not linear the most likely cause is
that the converter needs replacing. See Section 7.1.7 on NO
2
converter efficiency.
8.4. Calibration Frequency
To ensure accurate measurements of the NO, NO
X
, and NO
2
concentrations, calibrate the
analyzer at the time of installation, and re-calibrate it:
•
No later than three months after the most recent calibration or performance audit
which indicated analyzer calibration to be acceptable.
•
An interruption of more than a few days in analyzer operation.
•
Any repairs which might affect its calibration.
•
Physical relocation of the analyzer.
•
Any other indication (including excessive zero or span drift) of possible significant
inaccuracy of the analyzer.
Following any of the activities listed above, the zero and span should be checked to
determine if a calibration is necessary. If the analyzer zero and span drifts exceed the
calibration limits in Section 12 of the Q.A. Handbook
6
, a calibration should be performed.
8.5. Other Quality Assurance Procedures
Precision is determined by a one-point check at least once every two weeks. Accuracy is
determined by a three-point audit once each quarter.
044100102 Rev A
135
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...