Model 200E Instruction Manual
Calibration Procedures
If the
ZERO
or
SPAN
keys are not displayed during zero or span calibration, the measured
concentration value is too different from the expected value and the analyzer does not allow
to zero or span the instrument. Consult Section 11.3 for calibration problems.
7.3. Calibration Checks
Informal calibration checks, which only evaluate but do not alter the analyzer’s response
curve, are recommended as a regular maintenance item and in order to monitor the
analyzer’s performance. To carry out a calibration check rather than a full calibration, follow
these steps.
Connect the sources of zero air and span gas as shown in Figure 7-2, then perform the
zero/span calibration check procedure:
SAMPLE
NOX STB=XXX.X PPB
NO2=X.XX
< TST TST >
CAL
SETUP
ACTION:
Record the NO
X
,
NO
2
& NO
readings.
ACTION:
Supply span gas to the instrument
SAMPLE
NOX STB=XXX.X PPB
NO =X.XX
< TST TST >
CAL
SETUP
ACTION:
Record the NO
X
,
NO
2
& NO
readings.
Wait until
NOX
STB
is below
0.5 ppb. This
may take
several minutes.
The value of
NOX
STB
may jump
significantly.
Wait until it falls
below
0.5 ppb. This
may take several
minutes.
SAMPLE
NOX STB=XXX.X PPB
NO = X.X
< TST TST >
CAL
SETUP
The
SPAN
key appears during the transition from zero to
span. You may see both keys.
SAMPLE RANGE = 500.0 PPB
NOX=X.XX
< TST TST >
CAL
SETUP
Scroll the display to the
NOX STB
test function.
ACTION:
Supply the instrument with zero gas.
7.4. Calibration with Zero/Span Valves
To perform a manual calibration or calibration check of the analyzer with a zero/span valve
option installed and functioning, use the following method. The basic differences between
calibrating an M200E with the zero/span valve option instead of the basic calibration
method (Section 7.2) are that zero air and span gas are now supplied through dedicated
044100102 Rev A
115
Содержание 200E
Страница 15: ...M200E Documentation Model 200E Instruction Manual USER NOTES xiv 044100102 Rev A ...
Страница 19: ...M200E Documentation Model 200E Instruction Manual USER NOTES 4 044100102 Rev A ...
Страница 47: ...Frequently Asked Questions Glossary Model 200E Instruction Manual USER NOTES 32 044100102 Rev A ...
Страница 141: ...Calibration Procedures Model 200E Instruction Manual USER NOTES 126 044100102 Rev A ...
Страница 203: ...Theory of Operation Model 200E Instruction Manual USER NOTES 188 044100102 Rev A ...
Страница 239: ...Troubleshooting Repair Model 200E Instruction Manual USER NOTES 224 044100102 Rev A ...
Страница 243: ...APPENDIX A Model 200E Instruction Manual Figure A 3 Primary Setup Menu Except iDAS 228 044100110 Rev A ...
Страница 244: ...Model 200E Instruction Manual APPENDIX A Figure A 4 Secondary Setup Menu COMM VARS 044100110 Rev A 229 ...
Страница 245: ...APPENDIX A Model 200E Instruction Manual Figure A 5 Secondary Setup Menu DIAG 230 044100110 Rev A ...
Страница 246: ...Model 200E Instruction Manual APPENDIX A Figure A 6 Internal Data Acquisition iDAS Menu 044100110 Rev A 231 ...
Страница 265: ...APPENDIX A Model 200E Instruction Manual USER NOTES 250 044100110 Rev A ...
Страница 271: ...APPENDIX B Model 200E Instruction Manual USER NOTES 256 044100120 Rev A ...
Страница 275: ...APPENDIX D Model 200E Instruction Manual 260 044100141 Rev A USER NOTES ...