Teledyne API Model 200A NO
X
Analyzer Instruction Manual, 02246, Rev. F1
9.3.8
x
Sensor Module
Figure 9-9 and Figure 9-10 show the assembly of the NO
x
sensor module.
9.3.8.1 PMT
The PMT detects the light emitted by the reaction of NO with ozone. It has a gain of about
500,000 to 1,000,000. It is not possible to test the detector outside of the instrument in the field.
The best way to determine if the PMT is working is by using Optic Test. OT operation is
described in Section 9.1.3.3.
The basic method to diagnose a PMT fault is to eliminate the other components using ET, OT
and specific tests for other sub-assemblies.
9.3.8.2 Reaction Cell Temp
The reaction cell temperature is controlled by the CPU. It operates by reading a thermistor
amplifier on the 01086 Status/Temp board. The CPU controls the temperature by toggling a bit
on the V/F board. The V/F board TTL logic controls a solid state switch on the Switch Board in
the PSM. The switched 115VAC comes out of the PSM to a connector near the underside of the
reaction cell. A warning message may be present during initial warm-up or if the connector is
not plugged in after cleaning the reaction cell.
9.3.8.3 Preamp Board
The NO
x
Preamp Board is a multifunction board providing circuitry to support the following
functions.
1.
Preamp, buffer amplifier, physical range control hardware for the PMT detector.
2.
Precision voltage reference and voltage generation, and control for the PMT - HVPS
inside the sensor module.
3.
Constant current generator and adjustment for the Optic Test LED.
4.
Voltage generation and adjustment for Electric Test.
5.
Thermistor amplifier, control signal generation for the PMT cooler.
The setup and adjustment of items 1-4 above is covered in the Factory Calibration procedure in
Section 9.1.6. Item 5 has no adjustable features.
NO
9-62
Содержание 200A
Страница 10: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK x ...
Страница 12: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 1 2 ...
Страница 16: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 2 2 Rear Panel 2 4 ...
Страница 17: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 2 3 Inlet and Exhaust Venting 2 5 ...
Страница 22: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 2 4 Front Panel 2 10 ...
Страница 23: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 2 5 Assembly Layout 2 11 ...
Страница 26: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 2 14 ...
Страница 32: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 4 1 Block Diagram 4 2 ...
Страница 38: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 4 2 External Pump Pack 4 8 ...
Страница 40: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 4 10 ...
Страница 42: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 5 1 Sample Menu Tree 5 2 ...
Страница 43: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 5 2 Setup Menu Tree 5 3 ...
Страница 91: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 7 1 Calibration Setup 7 3 ...
Страница 109: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 7 2 Diagram of GPT Calibration System 7 21 ...
Страница 126: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 7 38 ...
Страница 130: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 8 1 Replacing the Particulate Filter 8 4 ...
Страница 134: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 8 2 Molybdenum Converter Assembly 8 8 ...
Страница 140: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 8 6 Pneumatic Diagram with IZS Option 8 14 ...
Страница 167: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 1 Span Calibration Voltage 9 25 ...
Страница 178: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 2 CPU Board Jumper Settings 9 36 ...
Страница 182: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 3 RS 232 PIN Assignments 9 40 ...
Страница 188: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 4 V F Board Settings 9 46 ...
Страница 193: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 5 Power Supply Module Layout 9 51 ...
Страница 194: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 6 Electrical Block Diagram 9 52 ...
Страница 198: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 7 Ozone Generator Subsystem 9 56 ...
Страница 201: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 8 Flow Pressure Sensor 9 59 ...
Страница 202: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 9 NOx Sensor Module 9 60 ...
Страница 203: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 10 NOx Sensor Module 9 61 ...
Страница 205: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 11 PMT Cooler Subsystem 9 63 ...
Страница 207: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 Figure 9 12 High Voltage Power Supply 9 65 ...
Страница 210: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 9 68 ...
Страница 218: ...Teledyne API Model 200A NOX Analyzer Instruction Manual 02246 Rev F1 INTENTIONALLY BLANK 10 8 ...