WorkHorse Rio Grande ADCP User's Guide
P/N 957-6167-00 (January 2001)
page 9
•
Place a light amount of DC-111 lubricant on the end-cap
connector pins (rubber portion only).
This will make it eas-
ier to connect or remove the I/O cable and dummy plug.
•
Do not connect or disconnect the I/O cable with power ap-
plied. An exception to this is the external battery case. The
external battery case connector is always “hot” when batter-
ies are installed. When you connect the cable with power
applied, you may see a small spark.
The connector pins
may become pitted and worn.
•
The WorkHorse I/O cable is
wet
mate-able, not
under water
mate-able.
4.3 Deployment
Guidelines
•
Align the compass whenever the battery pack or recorder
module is replaced, or when any ferrous metals are relocated
inside or around the WorkHorse housing.
Ferro-magnetic
materials affect the compass.
•
The AC power adapter is not designed to withstand water.
Use caution when using on decks in wet conditions.
•
Avoid using ferro-magnetic materials in the mounting
fixtures or near the Workhorse.
Ferro-magnetic materials
affect the compass.