WorkHorse Rio Grande ADCP User's Guide 

page 10 

RD Instruments 

Setup the WorkHorse Rio Grande ADCP 

Figure 5

 illustrates how to connect the WorkHorse cables and adapters on 

your workbench.  The internal battery plugs into a connector on the top 
circuit board.   
You will need a container of water large enough to submerge the Work-
Horse’s transducer head into during testing (two to three inches of water is 
sufficient).  Testing the WorkHorse out of water may cause some tests to 
fail but causes no harm to the WorkHorse.  

I/O CABLE

12 VDC BATTERY

COMPUTER

ADCP

 

Figure 5. 

WorkHorse Rio Grande Connections 

5.1 Serial 

Communication 

The standard communications settings for a WorkHorse Rio Grande is RS-
232, 9600-baud, no parity, 8 data bits and 1 stop bit.  If the serial protocol is 
set for RS232 and your computer expects RS422, you will need an RS232 
to RS422 adapter between the WorkHorse cable and your computer.  You 
can set the WorkHorse for baud rates other than 9600 baud.   

RS422

.  The WorkHorse Rio Grande is normally set for RS232, but it can be 

changed to RS422 by changing a switch setting.  The switch is in plain view 
on the top circuit board, near the cable connectors.  Its settings are plainly 
marked on the board.  This user’s guide assumes that you use RS232.   

5.2 

What if the WorkHorse Does Not Respond 

If your WorkHorse does not respond, check the serial port, cables, and 
power.  If necessary, refer to the 

Troubleshooting

 

Guide

 in the WorkHorse 

technical manual. 

Содержание WorkHorse Rio Grande ADCP

Страница 1: ...WorkHorse Rio Grande ADCP User s Guide P N 957 6167 00 January 2001 RD Instruments Acoustic Doppler Solutions...

Страница 2: ......

Страница 3: ...llation 11 6 3 Utility Software 12 7 Power 13 7 1 Bench Top Battery Power Requirements 13 7 2 Operation Modes 13 8 Testing Your WorkHorse 14 9 Compass Calibration 15 9 1 Preparing for Calibration 15 9...

Страница 4: ...orkHorse Rio Grande Connections 10 Figure 6 Compass Alignment 17 Figure 7 Visual Inspection before Deployment 19 Figure 8 Prepare the ADCP 20 Figure 9 Real Time Deployment 21 List of Tables Table 1 Wo...

Страница 5: ...Technical Manual guides Where needed there are references to detailed information and fig ures contained in the WorkHorse Technical Manual WorkHorse Rio Grande deployments are most often Real Time Rea...

Страница 6: ...nd channel discharge Plume tracking Suspended sediment load estimation Environmental surveys Plume tracking Planning new ports Bridge scouring Current mapping Simultaneous bathymetry discharge flow st...

Страница 7: ...ramics Never set the transducer on a hard surface The urethane face may be damaged Housing The standard WorkHorse housing allows deployment depths to 200 meters Thermistor The Thermistor measures the...

Страница 8: ...Horse Rio Grande ADCP User s Guide page 4 RD Instruments URETHANE FACE BEAM 3 MARK THERMISTOR PRESSURE SENSOR OPTIONAL TRANSDUCER HEAD HOUSING END CAP I O CABLE CONNECTOR Figure 1 Rio Grande ADCP Over...

Страница 9: ...on the end cap connector as the I O cable is being disconnected Stressing the end cap connector may cause the ADCP to flood Read the Maintenance guide for details on disconnecting the I O cable Figure...

Страница 10: ...7 are available in 16 20 40 85 and 220 MB cards The internal recorder holds two cards for a maximum of 440 MB of recording space The PC Card recorder is located on the Digital Signal Processor DSP bo...

Страница 11: ...page 7 Figure 4 Memory Card Overview 3 4 Spare Parts The following parts are included in the spare parts kit Table 2 Spare Parts Description Part number O ring face 2 260 Desiccant sealed bag DES3 Lu...

Страница 12: ...ns ducer faces on the Workhorse if it will be exposed to sun light Do not expose the I O connector to prolonged sunlight The plastic may become brittle Cover the connector on the Workhorse if it will...

Страница 13: ...you connect the cable with power applied you may see a small spark The connector pins may become pitted and worn The WorkHorse I O cable is wet mate able not under water mate able 4 3 Deployment Guid...

Страница 14: ...standard communications settings for a WorkHorse Rio Grande is RS 232 9600 baud no parity 8 data bits and 1 stop bit If the serial protocol is set for RS232 and your computer expects RS422 you will n...

Страница 15: ...to use WinADCP see the WinADCP User s Guide Documentation CD The Documentation CD has an Adobe Acrobat pdf elec tronic version of the WorkHorse Technical Manual Use the Documentation CD to search for...

Страница 16: ...I comma delimited format data described by a decoder DEC file into the raw BroadBand ADCP data file infilename result ing in an output ADCP data file called outfilename See BCONV DOC for decoder file...

Страница 17: ...se is asleep it consumes less than one mw A standard battery pack supplies sleep power for years At every opportunity the WorkHorse will sleep to conserve power while de ployed 7 2 Operation Modes The...

Страница 18: ...the tests DumbTerm runs gives you guidelines for running the tests and tells you what the results mean Table 5 WorkHorse ADCP Tests Test Guidelines Results PS0 Displays system parameters Verify the i...

Страница 19: ...orse on a piece of strong cardboard on top of a smooth wooden non magnetic table If a wooden table is not available place the WorkHorse on the floor as far away from metal objects as possible Use the...

Страница 20: ...vious cali bration While you are turning the WorkHorse for the two calibration rota tions the WorkHorse checks the quality of the previous calibration and dis plays the results It compares these resul...

Страница 21: ...the calibration procedure is not successful return your WorkHorse to the original factory calibration by using the AR command Try using the AR command if you have trouble calibrating your compass In s...

Страница 22: ...fouling The sensor port is surrounded by the antifouling paint but bio fouling may build up on the screw and eventually clog the sensor port However most organ isms do not seem to find the small amoun...

Страница 23: ...dition All mounting hardware installed Transducer faces clean and free from defects Verify compass alignment using DumbTerm if necessary re calibrate Are biofouling precautions needed Zero pressure se...

Страница 24: ...t specified WorkHorse deployment depths and temperatures Remember that desiccant rapidly absorbs moisture from normal room air Replace the desic cant whenever the WorkHorse housing or end cap is re mo...

Страница 25: ...s The Workhorse corrects data for tilts as large as 15 but tilts reduce the effective range and increase the depth of surface contamination Anti fouling paint You are free to use any anti fouling pain...

Страница 26: ...l to the average sound speed if the average sound speed is 1 less than the sound speed at the transducer the distance to the depth cell is 1 less than the displayed distance 12 2 When was the Data The...

Страница 27: ...angles the WorkHorse maps depth cells to other cells at the same depth If you want to make measurements near the surface from a bottom mounted WorkHorse you should minimize the tilt Depth cells are m...

Страница 28: ...group in any of the following ways RD Instruments RD Instruments Europe 9855 Businesspark Ave 5 Avenue Hector Pintus San Diego California 92131 06610 La Gaude France 858 693 1178 33 0 492 110 930 FAX...

Отзывы: