CU - Set Update Mode
Purpose
Configures the Sentinel V Real-Time ADCP’s WLAN.
Format
CUx
Range
x = 0 = OFF, 1 = Wireless, 2 = Ethernet (debug)
Default
CU0
Recommended Setting. Use as needed.
Description Send the CU1 command to enable the Sentinel V Real-Time ADCP’s WLAN. Wait for the
short beep and then connect to the ADCP’s WLAN and then enter 192.168.0.2 in the
browser address bar. This will open the Sentinel-V Utilities page, which allows for firmware
updates and downloading logs.
>cu1
Configuring network interface. Please wait
Use http://SV17302.adcp to access update link
You can also use http://192.168.0.2 to access update link
>
>cu2
Configuring network interface. Please wait
Network configured. Please connect to SV17302 Ethernet network.
Use http://SV17302 to access update link
You may need to refresh your browser. Hit CTRL F5
You can also use http://10.20.229.11 to access update link.
>
The CU command is not needed for Ethernet systems. Access the Sentinel-V Utilities page
using Sentinel V RT Utilities software.
Figure 96.
Sentinel-V Utilities Page
For information on Firmware Updates and Download Logs, see the Sentinel V_SC_RT