WaveSurfer 3000 Oscilloscopes
WaveSource Automatic Waveform Generator
The WaveSource Automatic Waveform Generator allows you to output custom sine, square, triangle, pulse, DC and noise
waveforms from the WaveSurfer 3000.
Connect a BNC cable from the WaveSource output on the back of the oscilloscope to the input device.
Choose
Utilities > WaveSource
or touch the front panel
WaveSource button
.
Check
Enable
, then choose a
Waveform
type.
Depending on the type, custom waveforms may be configured using different groups of
Settings
(e.g., Frequency or
Period, but not both). When you see buttons offering this choice, first select the group of settings to use. Adjust default
parameters by touching the field and turning the front panel Adjust knob, or touch twice and use the soft keypad to enter
new values.
Default Setup
returns the waveform settings to the defaults.
Calibrate
performs the internal calibration of the Function Generator output.
CAUTION.
Remove all cables before starting the calibration.