Operator’s 

Manual

 

PP009 Passive Probe 

Содержание PP009

Страница 1: ...PP009 1 PP009 2 Operator s Manual PP009 Passive Probe ...

Страница 2: ...yne LeCroy at its option will either repair or replace any assembly returned within its warranty period to the Customer Service Department or an authorized service center However this will be done only if the product is determined by Teledyne LeCroy s examination to be defective due to workmanship or materials and the defect is not caused by misuse neglect accident abnormal conditions of operation...

Страница 3: ...derstood and met DOUBLE INSULATION PROTECTIVE EARTH TERMINAL Precautions To avoid personal injury or damage to property review and comply with the following safety precautions Use product only as specified Connect only to grounded instruments Use only with compatible Teledyne LeCroy oscilloscopes that have their BNC input connected to an earth ground Do not connect the probe reference lead to any ...

Страница 4: ... s case or touching exposed connections may result in electric shock Use only within operational environment listed Use indoors only Do not use in wet or explosive atmospheres Keep product surfaces clean and dry Use only accessories compatible with the probe Handle with care The probe tip is extremely sharp and may puncture skin or cause other bodily injury if not handled properly Keep fingers beh...

Страница 5: ...only Do not operate where conductive pollutants may be present Maximum input voltage derates when frequency response is approximately 30 kHz or higher See voltage derating curve p 4 Specifications Electrical Characteristics Attenuation 10 1 Input Resistance 10 MΩ 1 Input Capacitance 9 5 pF Input Impedance see plot on p 5 Compensation Range 10 20 pF Bandwidth 500 MHz 3 dB Electrical Ratings Max Inp...

Страница 6: ... to 71o C Altitude Operating Up to 2000 m 6560 ft Humidity Operating 5 to 85 RH to 30 C 5 to 65 RH 30 C to 40 C 5 to 45 RH above 40 C General Characteristics Cable Length 1 3 m 4 3 ft Weight probe only 46 g Max VIN vs Frequency Frequency Hz 100 K 1 M 100 M 10 M 0 VIN Vrms 50 150 250 350 400 300 200 100 ...

Страница 7: ...ssories are shown with the Teledyne LeCroy part number followed by the description The following kits containing an assortment of probe connectivity accessories can be ordered directly from Teledyne Lecroy PKIT1 5MM 102 Basic Adapter Kit replaces the standard accessories with Qty 1 each of PK1 5MM 110 PK1 5MM 104 PK1 5MM 103 PK1 5MM 108 PK1 5MM 117 PK1 5MM 101 PK1 5MM 102 PK1 5MM 118 and PK1 5MM 1...

Страница 8: ...106 Color Coding Rings set Protective Cap Optional Accessories GROUND LEADS PK1 5MM 119 Ground Lead with mini clip PK1 5MM 120 Ground Lead 8 mm socket PK1 5MM 121 Ground Lead 2 mm plug PK1 5MM 122 Ground Lead 4 mm plug PK1 5MM 123 High Frequency Compensated Ground Lead The PK1 5MM 123 High Frequency compensated ground lead allows operation with long ground lead with minimum signal distortion ...

Страница 9: ...nd Spring INPUT ADAPTERS AND CLIPS PK1 5MM 111 Single Lead Adapter PK1 5MM 112 Dual Lead Adapter PK1 5MM 116 Adapter 2 mm plug PK1 5MM 117 Adapter 4 mm plug PK1 5MM 113 Pico Hook PK1 5MM 114Y PK1 5MM 114G Micro Clip Long PK1 5MM 115Y PK1 5MM 115G Micro Clip Short PK1 5MM 107 PCB Adapter PK1 5MM 110 BNC Adapter For low voltage use only 42 Vpk AC DC ...

Страница 10: ...ther maintenance and component replacement should be referred to qualified personnel Cleaning The outside of the probe should cleaned with a soft cloth dampened with either deionized distilled water or isopropyl alcohol Allow the surface to dry completely before returning the probe to service Never immerse the probe in any liquid Use of PCB Adapter The PCB adapter Teledyne Lecroy P N PK1 5MM 107 i...

Страница 11: ...for LF compensation is performed by connecting the input of the probe to a low frequency square wave such as the oscilloscope calibrator signal set to 1 kHz The compensation is adjusted by rotating the adjustment accessible through the small hole in the center of the housing near the BNC connector Use the tool supplied with the probe for this adjustment Undershoot Overshoot Correct adjustment Shou...

Страница 12: ...ct the proper tip for your application needs To replace the spring loaded tips carefully grip the outermost portion of tip and pull straight out along the axis of the probe using needle nose pliers Do not grip the plastic insulator with pliers this will squeeze the tip making it difficult or impossible to remove Do not grip the outer gold plated tube that the tip slides into With the tip removed a...

Страница 13: ...cted to utilization points socket outlets and similar of the low voltage mains installation Pollution Degree 2 operating environment where normally only dry non conductive pollution occurs Conductivity caused by temporary condensation should be expected Environmental Compliance END OF LIFE HANDLING The probe is marked with this symbol to indicate that it complies with the applicable European Union...

Страница 14: ...PP009 Passive Probe 12 ...