![TELEDYNE API T300 Скачать руководство пользователя страница 63](http://html.mh-extra.com/html/teledyne-api/t300/t300_operation-manual_1080880063.webp)
Teledyne API – Model T300/T300M CO Analyzer
Getting Started
61
•
At least 0.2m long;
•
No more than 2m long and;
•
Vented outside the shelter or immediate area surrounding the instrument.
A similar vent line should be connected to the VENT SPAN outlet on the back of the
analyzer.
E
XHAUST
O
UTLET
Attach an exhaust line to the analyzer’s EXHAUST outlet fitting. The exhaust line
should be:
•
PTEF tubing; minimum O.D ¼”;
•
A maximum of 10 meters long;
•
Vented outside the analyzer’s enclosure.
3.3.2.6.
PNEUMATIC LAYOUT FOR AMBIENT ZERO/PRESSURIZED SPAN OPTION
INSTRUMENT CHASSIS
Flow / Pressure
Sensor PCA
SAMPLE
PRESSURE
SENSOR
FLOW
SENSOR
Sample Gas
Flow Control
SAMPLE
GAS INLET
EXHAUST
GAS OUTLET
Particulate
Filter
PUMP
ZERO AIR
INLET
PRESSURE
SPAN INLET
VENT SPAN
OUTLET
Shutoff
Valve
Sample / Cal
Valve
Zero / Span
Valve
GFC Wheel
Housing
GFC Motor Heat
Sync
SAMPLE CHAMBER
Zero
Gas
In
Sample
Gas
In
Span
Gas
In
Figure 3-24:
Internal Pneumatic Flow OPT 50B – Zero/Span/Shutoff Valves
Table 3-9: Zero/Span Valve Operating States for Option 50B
MODE
VALVE
CONDITION
SAMPLE
(Normal
State)
Sample/Cal
Open to SAMPLE inlet
Zero/Span
Open to IZS inlet
Shutoff Valve
Closed
ZERO CAL
Sample/Cal
Open to ZERO/SPAN valve
Zero/Span
Open to IZS inlet
Shutoff Valve
Closed
SPAN CAL
Sample/Cal
Open to ZERO/SPAN valve
Zero/Span
Open to SHUTOFF valve
Shutoff Valve
Open to PRESSURE SPAN Inlet
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...