Calibration Procedures
Teledyne API – Model T300/T300M CO Analyzer
196
9.2.3.2.
ZERO/SPAN POINT CALIBRATION PROCEDURE
Only appears if either the
O
2
or the CO
2
sensor is
installed.
Only appears if either the
O2 or CO2 Sensors are
installed.
Press
ENTR
to changes
the
OFFSET
&
SLOPE
values for the CO
measurement.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE
GAS TO CAL:CO
CO
CO2
O2
ENTR
EXIT
SAMPLE
STABIL= XXXX PPM
CO=XXXX
< TST TST > CAL MSG SETUP
M-P CAL
STABIL= XXXX PPM
CO=XXXX
<TST TST>
ZERO
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
EXIT
at this point
returns to the
SAMPLE menu.
Press
ENTR
to change
the
OFFSET
&
SLOPE
values for the CO
measurement.
Press
EXIT
to leave the
calibration unchanged and
return to the previous
menu.
SAMPLE RANGE=50.0PPM CO=XX.XX
< TST
TST >
CAL MSG SETUP
Set the Display to show
the
STABIL
test function.
This function calculates
the stability of the CO
measurement.
Toggle
TST>
button until ...
Allow zero gas to enter the sample port
at the rear of the analyzer.
Wait until
STABIL
falls
below
0.2 PPM (for
T300) or 1.0 PPM (for
T300M).
This may take several
minutes.
SAMPLE
STABIL= XXXX PPM
CO=XXXX
< TST TST >
CAL
SETUP
M-P CAL
STABIL= XXXX PPM
CO=X.XXX
<TST TST>
ENTR
CONC
EXIT
SAMPLE
GAS TO CAL:CO
CO
O2
ENTR
EXIT
M-P CAL
STABIL= XXXX PPM
CO=X.XXX
<TST TST> ZERO
SPAN
CONC
EXIT
SAMPLE
RANGE TO CAL:LOW
LOW
HIGH
ENTR
EXIT
Allow span gas to enter the sample port
at the rear of the analyzer.
Wait until
STABIL
falls
below
0.2 PPM (for
T300) or 1.0 PPM (for
T300M).
This may take several
minutes.
SAMPLE
STABIL= XXXX PPM
CO=XXXX
< TST TST >
CAL
SETUP
M-P CAL
STABIL= XXXX PPM
CO=X.XXX
<TST TST>
ENTR
CONC
EXIT
M-P CAL
STABIL= XXXX PPM
CO=X.XXX
<TST TST> ENTR
CONC
EXIT
The
SPAN
button now
appears during the transition
from zero to span.
You may see both buttons.
If either the
ZERO
or
SPAN
buttons fail to appear see
Section 11 for
troubleshooting tips.
On instruments with a O
2
or CO
2
sensor installed
the analyzer will
simultaneously display the
gas measurements
throughout this procedure.
Only appears if the
DUAL
or
AUTO
range modes are
selected.
Use these buttons to
choose the appropriate
range.
Repeat entire procedure
for each range.
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...