Teledyne API – Model T300/T300M CO Analyzer
Setup Menu
131
5.10.1.
SETTING THE T300 CONCENTRATION ALARM LIMITS
To enable either of the CO concentration alarms and set the limit points, press:
SETUP X.X
CO ALRM 2: OFF
ON
ENTR EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
ALRM
EXIT
DIAG FCAL CO ALARM 2=300.00 PPM
3
0
0
.0
0
ENTR
EXIT
Toggle these
buttons to Set
the alarm point
EXIT
discards the new
setting
ENTR
accepts the
new setting
SAMPLE
RANGE=50.0 PPM CO= XX.XX
<TST
TST> CAL
SETUP
SETUP X.X
CO ALRM 2, DISALBED
NEXT
EDIT
PRNT EXIT
Toggle this button to
enable/disable
the alarm
Continue pressing
NEXT
until the desired
Alarm is selected
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...