Setup Menu
Teledyne API – Model T300/T300M CO Analyzer
122
An alternative method for measuring the output of the Current Loop converter is to
connect a 250 ohm
±
1% resistor across the current loop output in lieu of the current
meter (see Figure 3-9 for pin assignments and diagram of the analog output connector).
This allows the use of a voltmeter connected across the resistor to measure converter
output as VDC or mVDC.
V
+DC Gnd
Recording
Device
V IN +
V IN -
ANALYZER
V OUT +
V OUT -
250
Ω
Volt
Meter
Figure 5-6:
Alternative Setup Using 250Ω Resistor for Checking Current Output
Signal Levels
In this case, follow the procedure above but adjust the output for the following values:
Table 5-8:
Current Loop Output Check
% FS
Voltage across Resistor for
2-20 mA
Voltage across
Resistor for 4-20 mA
0
500 mVDC
1000 mVDC
100
5000 mVDC
5000 mVDC
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...