Teledyne API – Model T300/T300M CO Analyzer
Setup Menu
117
5.9.3.5.
INDIVIDUAL CALIBRATION OF THE ANALOG OUTPUTS
To use the
AUTO CAL
feature to initiate an automatic calibration for an individual
analog output, select the
ANALOG I/O CONFIGURATION
submenu (see Figure 5-3)
then press:
From the
AIO CONFIGURATION SUBMENU
DIAG AIO CONC_OUT_2: 5V, CONC2, NOCAL
<SET
SET>
EDIT
EXIT
Continue pressing
SET>
until ...
DIAG AIO CONC_OUT_2: CALIBRATED:NO
<SET
SET>
CAL
EXIT
DIAG AIO CONC_OUT_2: RANGE: 5V
SET>
EDIT
EXIT
DIAG AIO
AUTO CALIBRATING CONC_OUT_2
DIAG AIO CONC_OUT_2: CALIBRATED: YES
<SET
SET> CAL
EXIT
DIAG
ANALOG I/O CONFIGURATION
PREV NEXT
ENTR
EXIT
Continue pressing
SET>
until you reach the
output to be configured
DIAG AIO AOUTS CALIBRATED: NO
SET>
CAL
EXIT
06864D DCN7562
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864D DCN7562 ...
Страница 80: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 78 This page intentionally left 06864D DCN7562 ...
Страница 134: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 132 This page intentionally left blank 06864D DCN7562 ...
Страница 182: ...06864D DCN7562 ...
Страница 190: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 188 This page intentionally left blank 06864D DCN7562 ...
Страница 227: ...225 This page intentionally left blank 06864D DCN7562 ...
Страница 228: ...06864D DCN7562 ...
Страница 320: ...INDEX Teledyne API Model T300 T300M CO Analyzer 318 This page intentionally left blank 06864D DCN7562 ...
Страница 347: ...Appendix B Schematic 06864D DCN7562 ...