![TELEDYNE API T108 Series Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/teledyne-api/t108-series/t108-series_addendum-to-operational-manual_1080878015.webp)
Teledyne API – Model T108 Addendum to T100 Manual
Introduction
07268C DCN8258
13
Figure 1-5. T108-to-501TS Rear Panel Pneumatic Connections
1.6. OPERATION AND CALIBRATION
Refer to the T100 manual for the overall operation of the SO
2
analyzer. This unit has some unique
operating characteristics and calibration procedures detailed below.
The basic purpose of this instrument is to analyze CO
2
sample gas for sulfur containing impurities.
Typically, the impurities should be at low levels; therefore, it is especially important that the zero
calibration of the analyzer is done accurately so that even small levels of impurities can be detected.