8.2.1. SENSOR OK 

The Sensor OK LED indicates the status of the O

3

 sensor module in the monitor. 

The normal state of this LED (or Status Output) is On. If this LED remains off after 

the normal warm-up period, then a failure has occurred and the monitor should be 

removed from service and repaired. 
The most common cause of this warning is a failure of the UV LED. If the UV LED 

intensity has dropped to a level that cannot meet the specified values, then it should 

be replaced (See Section 7.5). 

8.2.2. INVALID READING 

The Invalid  Reading  LED indicates that  the instrument is  reading  a value that 

cannot be  represented properly on the analog output.  Since the analog output is 

limited to 0-5V (or 4-20mA), it cannot properly represent negative values, or values 

in excess of the full-scale range. 
If the monitor is consistently reading negative values, then a zero calibration should 

be performed, see Section 5.2.3.2. 
If the monitor is consistently reading values in excess of the full-scale range, then 

the range value should be adjusted higher. See Section 5.2.4 for details on setting 

the range value. 

8.2.3. CHECK LAMP 

The Check Lamp status indicates that the UV LED intensity has dropped to a level 

where the UV LED needs to be replaced (Section 7.5) the Sensor OK LED will also 

turn off, indicating that the monitor must be immediately serviced or taken off-line.  

8.2.4. PNEUMATIC ERROR 

The Pneumatic Error LED indicates that one of the pneumatic parameters, flow or 

pressure, has gone outside of normal ranges. Note that this is a non-critical warning. 

Unless it is also accompanied by Sensor OK LED turning off, the instrument does 

not need to be removed from service at this time. 
The first step in troubleshooting a Pneumatic Error is determining which parameter 

has  caused the warning.  At the monitor front panel, navigate to  the VIEW menu 

(See Section 5.2.2). Examine the Flow and Pressure values and compare them to the 

limits described in Table 8-1 and take appropriate action as described below. 

8.2.4.1. P

RESSURE 

T

OO 

H

IGH

 

 

The gas inlet pressure is too high. Verify that the ozone gas and zero gas 

delivery pressure is regulated to between 9 and 18 psig (high pressure error 

occurs when pressure exceeds 18 psig). 

 

Something is restricting the flow in the monitor. Check for restriction in the 

exhaust gas line.  Check for kinked or obstructed tubing inside the monitor. 

 

The pressure sensor has failed or drifted.   

Содержание Everywhereyoulook 480H

Страница 1: ...e API TAPI 9970 Carroll Canyon Road San Diego CA 92131 1106 Toll Free 1 800 324 5190 Phone 1 858 657 9800 Fax 1 858 657 9816 Email api sales teledyne com Website http www teledyne api com Copyright 20...

Страница 2: ...E IS INTENTIONALLY LEFT BLANK...

Страница 3: ...ecific information CAUTION Hot Surface Warning Do Not Touch Touching some parts of the instrument without protection or proper tools could result in damage to the part s and or the instrument Technici...

Страница 4: ...renseignements sp cifiques MISE EN GARDE Surface chaude Ne pas toucher Toucher certaines parties de l instrument sans protection ou sans les outils appropri s pourrait entra ner des dommages aux pi c...

Страница 5: ...ment provided but not manufactured by TAPI is warranted and will be repaired to the extent and according to the current terms and conditions of the respective equipment manufacturer s warranty Product...

Страница 6: ...iv Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 This page intentionally left blank...

Страница 7: ...NECTIONS 24 3 5 1 Installing Flow Control Assemblies 24 3 5 2 Ozone Inlet Connection 24 3 5 3 Zero Gas Inlet Connection 25 3 5 4 Exhaust Connection 25 3 6 PNEUMATIC FLOW DIAGRAM 25 4 CALIBRATION 26 4...

Страница 8: ...7 5 UV LED REPLACEMENT 46 7 6 SENSOR MODULE REPLACEMENT 47 7 7 DISPLAY ASSEMBLY REPLACEMENT 48 7 8 CLEANING EXTERIOR SURFACES OF THE 480H 49 8 TROUBLESHOOTING 50 8 1 REFERENCE DRAWINGS 50 8 2 TROUBLE...

Страница 9: ...ure 3 6 Pneumatic Connections 24 Figure 3 7 Pneumatic Diagram 25 Figure 5 1 Front Panel Display and Controls 27 Figure 5 2 Front Panel Selection and Control Buttons 28 Figure 6 1 RS 232 Signal Configu...

Страница 10: ...Status Output Definitions 22 Table 5 2 Sensor Menu Parameters 30 Table 5 3 Analog Step Test Values 30 Table 5 4 VARS List 32 Table 5 5 Alarm Configuration Settings 34 Table 6 1 RS 232 RS 485 Port Set...

Страница 11: ...e times with minimal zero drift for accurate and stable readings and it has built in tests and diagnostics to allow maximum uptime We at Teledyne API will be pleased to provide you with any support re...

Страница 12: ...10 Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 This page intentionally left blank...

Страница 13: ...ure Range 3 0 30 0 psig Sample Flow Rate 0 2 2 0 LPM Humidity Range 10 90 RH non condensing Temperature Range 5 45o C Dimensions H x W x D NEMA 4X 17 35 x 15 96 x 6 80 440 7 mm x 405 4 mm x 172 7 mm W...

Страница 14: ...61010 1 2010 3rd Edition Safety requirements for electrical equipment for measurement control and laboratory use 2 2 2 EMC IEC EN 61326 1 2010 Class A Emissions Industrial Immunity FCC 47 CFR Part 15B...

Страница 15: ...ion explosion hazard due to flammable gases vapors liquids never allow contaminants to enter monitor during installation or use The combination of ozone and the catalytic ozone destruct media in the m...

Страница 16: ...lation and for access to make adjust all connections The monitor must be mounted securely with four 4 bolts or anchors to a vertical structure or wall capable of supporting 80 lbs The mounting bolts o...

Страница 17: ...09421A DCN8343 Teledyne API Model 480H O3 Monitor User Manual 15...

Страница 18: ...d on the 3 pin AC connector Failure to do so may result in a shock hazard and malfunction of the monitor 3 3 1 WIRING REQUIREMENTS Use appropriate wiring rated for this type of equipment ensuring that...

Страница 19: ...erminal block and grounding lug on the chassis AC power connection to the monitor should be made with 12 14 AWG stranded copper wire connected to the monitor as follows Earth Ground green Connect the...

Страница 20: ...the 480L by applying power to the monitor The front panel display will light and display a Teledyne splash screen The home screen will display the ozone concentration and flow after the monitor begin...

Страница 21: ...put See Section 3 4 1 1 for details on verifying or changing the analog output hardware configuration The analog output is automatically scaled to the range of the instrument See Section 5 2 4 2 for i...

Страница 22: ...he monitor 2 Open front panel 3 Locate J15 and J17 connectors on the Mainboard PCA remove Relay Expansion board if installed as indicated in Figure 3 5 4 Set the desired output as shown in Figure 3 5...

Страница 23: ...e same connector as the Analog output and which mirror the state the values shown in the status area of the menu structure They are defined in Table 3 1 Electrically these outputs are optically isolat...

Страница 24: ...GS OR DATA Contact life may be dramatically shortened if inductive loads are driven without any provision for minimizing high voltage inductive kick that can occur Table 3 2 Relay Functions for Single...

Страница 25: ...communication with the monitor connect an Ethernet cable from the Ethernet interface connector see Figure 3 4 to an Ethernet port that is connected to a LAN Although the default setting is DHCP enabl...

Страница 26: ...nd Teflon Attach the Flow Control Assemblies by screwing them onto the threaded fittings of the ozone inlet and the zero gas inlet Hand tighten orifice assemblies until snug With a 9 16 open end wrenc...

Страница 27: ...r occupied area Exhaust gas from the monitor may contain dangerous levels of ozone Connect a OD vent line to the Exhaust fitting on the outside bottom panel This line must be vented outdoors or to an...

Страница 28: ...edyne API also recommends that the monitor be re calibrated once a year Teledyne API can provide NIST traceable calibration services at our factory or on site Please contact Technical Support for deta...

Страница 29: ...Teledyne splash screen The home screen will display the ozone concentration and flow after the monitor begins taking stable readings 5 1 FRONT PANEL OVERVIEW The front panel provides a display screen...

Страница 30: ...t accepting selected input Enter Accepts the selected input Pressing and holding any button for at least two seconds or pressing no buttons for at least 30 seconds returns the display directly to the...

Страница 31: ...w span gas to enter for 1 hour prior to execute Flow Cal set flow target based on external flow meter execute Alarm screen becomes Default Screen when alarm is triggered solid fill indicates which ala...

Страница 32: ...m up period 5 2 3 DIAG MENU The DIAG menu contains functions that are useful for testing and configuring external equipment that may be connected to the rear panel I O of the instrument 5 2 3 1 ANALOG...

Страница 33: ...ise or dithering of the concentration about the zero point may occur and is normal This noise is typically 0 5 of the full scale range in magnitude 5 2 3 3 CAL SPAN The Span Calibration also allows th...

Страница 34: ...Status Failed message will be displayed and you will need to return to the start of the Flow Cal menu If the Cal Status Failed message occurs it means the Flow Cal cannot be performed because it woul...

Страница 35: ...ges zero calibration ON ON OFF ANALOG_RANGE Full scale concentration range Used to scale the analog output 20 5 50 g Nm3 0 5 25 0 WT ANALOG HW CONF Determines if the analog output uses 4 20mA or 0 5V...

Страница 36: ...ly reset when the alarm condition is removed The Reset button has no effect when configured for alarms set to LATCHING OFF 5 2 4 1 SETUP COMM MENU The Setup Comm menu includes two submenus Comm Addr a...

Страница 37: ...API Model 480H O3 Monitor User Manual 35 5 2 5 STATUS SCREEN Provides the status of Pneumatic Sensor Lamp and Valid Reading functions and becomes the default screen any time an Error occurs See Secti...

Страница 38: ...nection use the front panel control keys Section 5 1 2 to navigate to the Setup Comm Ethernet IP mode menu If you don t know what the IP Address Mask or Gateway values should be consult with your IT A...

Страница 39: ...e To change or verify the configuration of the serial COM refer to Figure 6 1 while performing the following instructions 1 Disconnect power from the monitor 2 Open front panel 3 Locate J20 and J23 co...

Страница 40: ...38 Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 Figure 6 1 RS 232 Signal Configurations...

Страница 41: ...nction Code Description 01 Read Coils 02 Read Discrete Inputs 03 Read Holding Registers 04 Read Input Registers 05 Write Single Coil 06 Write Single Register 15 Write Multiple Coils 16 Write Multiple...

Страница 42: ...nverse Float Double 32 8 Sensor Gas Temperature K AI_CELL_TEMP Inverse Float Double 32 10 Stream 1 Concentration PPB STREAM_1_O3_CONC Inverse Float Double 32 Discrete Input Registers Read Only Registe...

Страница 43: ...EE 754 format read in highword low word order Register Address Description Tag Name Data Type Bits 12 Stream 1 HI Alarm Limit PPB STREAM_1_HI_ALARM_SETPOINT Inverse Float Double 32 24 Stream 1 HI HI A...

Страница 44: ...cause eye damage Always use safety glasses made from UV blocking material Generic plastic glasses are not adequate 7 1 MAINTENANCE SCHEDULE Table 7 1 below outlines the suggested maintenance procedur...

Страница 45: ...sent when power is connected to the instrument Figure 7 1 shows the internal layout of the 480H These figures will be referenced in the procedures that follow Note the caution areas Power Supply and E...

Страница 46: ...g and inspect for any cracks or deformities If there are cracks or deformities replace the o ring with a new one otherwise it can be reused 5 Remove the two sample filter retainers and glass fiber fil...

Страница 47: ...removed by sliding upwards 7 Remove the two 2 mounting screws using a 2 Phillips screwdriver Note that there are four screw heads visible on the top of the valve body only two 2 of these should be re...

Страница 48: ...sconnecting its power at the J12 connector Figure 7 3 and removing the two larger socket head screws Figure 7 4 to access the UV LED After installing and securing the replacement UV LED PCA assembly i...

Страница 49: ...from the sensor PCA that run to J13 Bench Supply and J10 Bench Serial Interface on the Mainboard and by disconnecting the pneumatic tubing at the sample filter and at the input to the scrubber Then re...

Страница 50: ...t matter 3 Free the display cable from the securing clips 4 Carefully remove the EMI core from the enclosure and remove any remaining adhesive from the door 5 Using a Star T10 screwdriver remove the s...

Страница 51: ...splay and control keys are functioning properly Figure 7 6 Display Assembly Removal 7 8 CLEANING EXTERIOR SURFACES OF THE 480H If necessary the front panel mask and keyboard of the 480H can be cleaned...

Страница 52: ...ANEL STATUS LED AND STATUS OUTPUTS Figure 8 1 Display and Control Panel The status LED on the front panel reflects the current operating status of the monitor and indicates fault conditions See Sectio...

Страница 53: ...s Output Status LED Label Normal Operating State Triggers Critical Warning 1 Sensor OK On Reference 125 0 mV OR Reference 3120 0 mV Measure 3120 0 mV Yes 2 Invalid Reading Off O3 Concentration too low...

Страница 54: ...setting the range value 8 2 3 CHECK LAMP The Check Lamp status indicates that the UV LED intensity has dropped to a level where the UV LED needs to be replaced Section 7 5 the Sensor OK LED will also...

Страница 55: ...nsor has failed or drifted Contact Teledyne API Technical Spport for assistance 8 2 5 TECHNICAL ASSISTANCE If this manual does not solve your problems technical assistance may be obtained from Teledyn...

Страница 56: ...filled with gas scrubbed to remove ozone The ratio of the intensity of light passing through the scrubbed gas to that of the sample forms a ratio I Io This ratio forms the basis for the calculation o...

Страница 57: ...ility of ozone to absorb 254 nm light Most current measurements place this value at 308 cm 1 atm 1 at STP The value of this number reflects the fact that ozone is a very efficient absorber of UV radia...

Отзывы: