5.2.4.2. S
ETUP
A
LARM
M
ENU
The ALARM menu allows the operator to view and edit properties of the
concentration alarms. The table below summarizes the alarm configuration settings.
The Alarm screen becomes the default screen any time an alarm is triggered.
Table 5-5. Alarm Configuration Settings
Property
Meaning
Default Value
Range
ALARM_MODE
Controls the latching behavior
of the concentration alarms.
When Latching is enabled,
the alarm persists until ‘Reset’
is pressed to clear the alarm*
ON
ON/OFF
HI ALARM
HI alarm trigger value
1.00 WT%
1.00 g/Nm3
0.02%-25.0WT %
0.02-400 g/Nm3
HIHI ALARM
HI-HI alarm trigger value
1.02 WT%,
1.02 g/Nm3
0.02%-25.0WT%
0.02-400 g/Nm3
ON/OFF HI
ALARM
HI alarm enable, enables HI
alarm operation
ON
ON/OFF
ON/OFF HI HI
ALARM
HI-HI alarm enable, enables
HI-HI alarm operation
ON
ON/OFF
*The alarm cannot be reset if the ozone reading for that channel, or all channels for the global alarms, has
not come down below the alarm set point. Non-latching alarms will automatically reset when the alarm
condition is removed. The Reset button has no effect when configured for alarms set to LATCHING OFF
5.2.4.1. S
ETUP
>C
OMM
M
ENU
The Setup>Comm menu includes two submenus: Comm Addr and Ethernet.
Select Comm Addr to set the MODBUS address; the Register Maps are presented in
Section 6.3.
Select Ethernet to view Info or to set the IP Mode to Static or DHCP (Section 6.1).
5.2.4.2. S
ETUP
>F
ACTORY
D
EFAULT
M
ENU
The Factory Default menu restores the factory calibration settings, thereby erasing
user-customized settings; therefore, it is strongly recommended that Factory Default
be used only when explicitly directed by Technical Support (
) who will then need to provide a software application to
reload the factory settings.
To proceed with the Setup>Factory Default menu, select Restore which will trigger
a confirmation screen to either continue by pressing the Enter control key or abort
by pressing the Back control key (Section 0). Once you start the Factory Default
Restore procedure, follow the instructions displayed, and do not unplug the
instrument until instructed to on the display. When the procedure is complete, and
the factory settings are loaded, a zero calibration is required after rebooting the
instrument.
Содержание Everywhereyoulook 480H
Страница 2: ...E IS INTENTIONALLY LEFT BLANK...
Страница 6: ...iv Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 This page intentionally left blank...
Страница 12: ...10 Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 This page intentionally left blank...
Страница 17: ...09421A DCN8343 Teledyne API Model 480H O3 Monitor User Manual 15...
Страница 40: ...38 Teledyne API Model 480H O3 Monitor User Manual 09421A DCN8343 Figure 6 1 RS 232 Signal Configurations...