09420A DCN8343
Teledyne API Model 480M O
3
Monitor User Manual
21
Figure 3-5. Mainboard – Analog Output Configuration
3.4.2. STATUS OUTPUTS
Refer to Figure 3-4 for the location of the status outputs, which are on the same
connector as the Analog output and which mirror the state the values shown in the
status area of the menu structure. They are defined in Table 3-1.
Electrically these outputs are optically isolated digital outputs that consist of open
collector transistors with a common ground. They can be used to interface with
devices that accept logic-level digital inputs, such as Programmable Logic
Controllers (PLCs).