AFA4000/E
VAV Airflow Controller / p.78
Figure 41: Maximum allowable displacement for a range of fume hood widths
1.
The area protected by the light beams should cover the open area under the sash, typically
500 mm (20") from its base.
The light curtain transmitter and receiver can be positioned above the bottom of the sash in
the fume hood if they are free from obstructions.
Where possible, make sure the light path is aligned flush to the bottom of the sash base.
Note: There is a 123 mm (4.84") gap from the top of the light curtain transmitter to the first
LED down and a 22 mm (0.87") gap from the bottom of the transmitter to the first
LED up.
Содержание AFA4000/E
Страница 1: ...AFA4000 E VAV Airflow Controller Installation and operating manual Issue 2...
Страница 57: ...AFA4000 E VAV Airflow Controller p 57 Figure 22 AFA4000 E inverter control connections...
Страница 58: ...AFA4000 E VAV Airflow Controller p 58 Figure 23 AFA4000 E damper control connections...
Страница 59: ...AFA4000 E VAV Airflow Controller p 59 Figure 24 AFA4000 E typical connection diagram...
Страница 60: ...AFA4000 E VAV Airflow Controller p 60 Figure 25 AFA4000 E AS typical connection diagram...
Страница 62: ...AFA4000 E VAV Airflow Controller p 62 Figure 27 SM7 airflow sensor dimensions...
Страница 71: ...AFA4000 E VAV Airflow Controller p 71 Figure 34 Side views of a rack and pinion drive typical installation...
Страница 96: ...AFA4000 E VAV Airflow Controller p 96 6 9 Auto Sash System Control Box general wiring diagram...
Страница 158: ...AFA4000 E VAV Airflow Controller p 158...
Страница 182: ...AFA4000 E VAV Airflow Controller p 182 Figure 79 Connection diagram with RS232 485 converter...