14.
Cleaning and maintenance
14.1
Cleaning
The Auto Sash Controller and AFA4000/E are generally maintenance free.
All parts fitted within the laboratory space may need occasional cleaning depending on the
environmental conditions and usage. Include the following:
Part
Method
Notes
Keypad
Use a dry cloth, optical lens or
screen wipe to remove dust, dirt,
grease or finger marks.
Do not remove the keypad whilst
the unit is powered up. Isolate the
power if the keypad is removed for
cleaning.
Personnel Sensor
Use a dry cloth, optical lens or
screen wipe to remove dust, dirt,
grease or finger marks.
Do not adjust the sensor position
whilst cleaning. If the sensor angle
is adjusted the user detection may
be compromised.
Light Curtains
Use a dry cloth, optical lens or
screen wipe to remove dust, dirt,
grease or finger marks.
The Light Curtain is calibrated to
detect glass objects. Make sure the
light curtain is kept free from
obstructions and kept clean.
Under Sash sensor
Use a dry cloth, optical lens or
screen wipe to remove dust, dirt,
grease or finger marks.
Do not adjust the sensor position
whilst cleaning. If the sensor angle
is adjusted the sensor beam will be
compromised.
Содержание AFA4000/E
Страница 1: ...AFA4000 E VAV Airflow Controller Installation and operating manual Issue 2...
Страница 57: ...AFA4000 E VAV Airflow Controller p 57 Figure 22 AFA4000 E inverter control connections...
Страница 58: ...AFA4000 E VAV Airflow Controller p 58 Figure 23 AFA4000 E damper control connections...
Страница 59: ...AFA4000 E VAV Airflow Controller p 59 Figure 24 AFA4000 E typical connection diagram...
Страница 60: ...AFA4000 E VAV Airflow Controller p 60 Figure 25 AFA4000 E AS typical connection diagram...
Страница 62: ...AFA4000 E VAV Airflow Controller p 62 Figure 27 SM7 airflow sensor dimensions...
Страница 71: ...AFA4000 E VAV Airflow Controller p 71 Figure 34 Side views of a rack and pinion drive typical installation...
Страница 96: ...AFA4000 E VAV Airflow Controller p 96 6 9 Auto Sash System Control Box general wiring diagram...
Страница 158: ...AFA4000 E VAV Airflow Controller p 158...
Страница 182: ...AFA4000 E VAV Airflow Controller p 182 Figure 79 Connection diagram with RS232 485 converter...