How to Guide 

           Audio Loudness Monitoring 

24     

www.tektronix.com\wfm8000

 

www.tektronix.com\wfm7100

 

www.tektronix.com\wfm5200

 

 

Function

 

Description

 

Meter Scaling

 

You can change the range and limits of the Audio bar display in the 
Meter Scaling menu. The meter scaling choices are:  

 Normal  

  Custom Height  

 Custom Offset  

  Graticule Step Size  

 

If you use the 

Normal

 settings, then you will see 70 dB of range for the 

bars. For digital input, the top of the scale will be DBFS, and for analog 
inputs it will be dBu.  
If you chose the 

Custom Height

 settings, then you can reduce or 

expand the range. If you choose the 

Custom Offset

 setting, then you 

can shift the displayed levels up and down. Together, these allow you to 
zoom in on one region or shrink the bars to cover a large range.  
If you select 

Graticule Step Size

, then you can select spacing between 

graticules and labels on the display. For example, at a value of 

6

 there 

will be a label at 0dB, -6dB, -12dB, and so on. 

 

Alarm Thresholds 

Audio Loudness configurable Alarm Thresholds are:  
 

 

Channel Loud

: The loudness level threshold for a single channel, 

audio source, above which the loudness value triggers an alarm 
condition. 

 

Program Loud

: The loudness level threshold for all channels used 

by the Loudness measurement, above which the loudness value 
triggers an alarm condition. This alarm will trigger an in-bar audio 
message of Too Loud. 

 

Program Quiet

: The loudness level threshold for all channels used 

by the Loudness measurement, below which the loudness value 
triggers an alarm condition. This alarm will trigger an in-bar audio 
message of Too Quiet. 

 

Target/DN Delta

: When the difference between the Target loudness 

level threshold and the current Dialnorm value exceed the user 
defined setting an alarm will be triggered. 

 

Target Loudness 

 

High 

 

Low 

The 

Target Loudness

 of an audio program is defined as the specified 

value for the Anchor Element (Dialog Level), established to facilitate 
content exchange from supplier to operator. 
 
The Anchor Element is the perceptual loudness of a reference point 
element around which other elements are balanced in producing the 
final mix of the content, or that a reasonable viewer would focus on 
when setting the volume control, as defined in ATSC standard A/85. 
The default value within the instrument is -24LKFS and is user 
selectable. 
 
High and Low thresholds are defined above and below the Target 
Loudness value.  
The default values are 2LU (+/-2LKFS) above and below the Target 
Loudness. 
 
For detailed information on the ATSC A/85 standard visit 

www.atsc.org

 
 

Содержание WFM5200 & WVR5200

Страница 1: ...How to Guide Audio Loudness Monitoring WFM8300 8200 WVR8300 8200 WFM6120 7020 7120 WVR6020 7020 7120 WFM5200 WVR5200 ...

Страница 2: ...ion Version 6 1 4 also supports Dolby Digital Plus when the DPE or DDP option is installed within the instrument WFM8300 8200 WVR8300 8200 Version 2 5 x This release version 2 5 x adds support for Audio Loudness measurements to the ITU R BS 1770 standard with support for ATSC A 85 recommended practice and EBU R128 Version 2 5 x also supports Dolby Digital Plus when the DPE or DDP option is install...

Страница 3: ...ge loudness for the various programs they are transmitting Loudness is a subjective measurement and is primarily a psychological correlation with the physical intensity of the audio signal Different people will perceive the loudness of the signal differently and therefore a variety of techniques have been used to characterize the overall loudness of the signal The ITU R BS 1770 standard algorithms...

Страница 4: ...ssing FULL again will toggle back to FlexVu On WFM WVR5200 series use the Thumbnail button to cycle through the tiles and push and hold the Thumbnail button to toggle between Full and four tile modes 3 Press and hold the Audio button to display the menu 4 Move up and down the menu using the Arrow Keys or General Knob to the Loudness Meter menu 5 Enter the sub menu by pressing SEL and navigate to t...

Страница 5: ... the Loudness measurements within the Loudness Settings menu To simplify the operation of the instrument Loudness Presets are available to quickly configure the unit to the appropriate standard recommendations 1 Press the Config button to enter the configuration menu 2 Move up and down the menu using the Arrow Keys or General Knob to the Loudness Settings menu 3 Press SEL to enter the sub menu and...

Страница 6: ...pressing the HELP button on the front of the instrument Elements of Audio Display with Loudness Measurements Figure 3 Audio Tile Display 1 Loudness Meter is enabled from within the Audio Tile menu and can be selected to show the Loudness for the program a channel pair or disabled and not displayed within the audio tile The Program Loudness PL is a summation of a number of audio channels The channe...

Страница 7: ...applied then some measurement results may be excluded from the loudness measurements 4 Dialnorm An AC 3 metadata parameter numerically equal to the absolute value of the Dialog Level carried in the AC 3 bit stream Valid values are 1 31 The value of 0 is reserved The values of 1 to 31 are interpreted as 1 to 31 This display shows the current metadata value decode from the Dolby stream or from the S...

Страница 8: ...udio channel is included in the Loudness summation Whereas a White channel label indicates that the audio channel is not included in the Loudness summation The summation of the number of channels can be configured by the user within the Loudness Settings menu 8 Audio Program number Specifies the audio program currently being used for the program loudness Within the Config menu the audio program ca...

Страница 9: ...les and push and hold the Thumbnail button to toggle between Full and four tile modes 3 Press and hold the Audio button to display the menu 4 Move up and down the menu using the Arrow Keys or General Knob to the Aux Display menu 5 Enter the sub menu by pressing SEL and navigate to the Loudness selection Figure 4 Figure 4 Loudness Readouts within Audio Display Figure 4 provides large number readout...

Страница 10: ... full screen pressing FULL again will toggle back to FlexVu On WFM WVR5200 series use the Thumbnail button to cycle through the tiles and push and hold the Thumbnail button to toggle between Full and four tile modes 3 Press and hold the Status button to display the menu 4 Move up and down the menu using the Arrow Keys or General Knob to the Display Type menu 5 Enter the sub menu by pressing SEL an...

Страница 11: ...TU R BS 1771 For audio loudness measurements a calculation for each channel and channel pairs is shown in the audio session display along with the Program Loudness for the short and infinite duration The Program Loudness measurement is based on a summation of a number of the audio channels and the duration of the measurement The information on the number of channels summed together is provided wit...

Страница 12: ... Configure the Loudness Session Display 1 Select one of the tiles 1 2 3 4 and press the Status button 2 Select FULL to make the display full screen pressing FULL again will toggle back to FlexVu On WFM WVR5200 series use the Thumbnail button to cycle through the tiles and push and hold the Thumbnail button to toggle between Full and four tile modes 3 Press and hold the Status button to display the...

Страница 13: ...ilable within the instrument A trend chart Figure 6 that shows a graph of the loudness measurements for the last 90 or 180 seconds and a bar chart that shows the maximum and minimum loudness value during each bar period and can show the loudness variations from 7 5 minutes to 30 hours instrument Figure 7 Figure 6 Loudness Session Trend Chart 90 Seconds ...

Страница 14: ...rting can be temporarily muted Graph Scale The horizontal scale of the chart can be changed from 0 90 and 180 seconds trend chart to 7 5 minute to a 30 hour bar chart Graph Gain The vertical gain of the chart can be changed between 0 60 8 40 and 12 36dB Save History to USB Allows downloading of all loudness session data to a memory stick inserted into the front of the waveform monitor Alternativel...

Страница 15: ...oudness Session chart continue to accumulate until you manually restart the data collection or cycle the power With the Loudness Session displayed and selected as the active tile Bright Blue border around tile Press and hold the STATUS button to display the STATUS pop up menu Scroll to History Reset and press SEL to reset the session Figure 7 Loudness Session Bar Chart six hour duration ...

Страница 16: ... based on duration of audio session 5 Dialnorm is the metadata value that is present either within the Dolby data stream or carried as an Ancillary data packet per SMPTE2020 6 True Peak is the maximum true peak value recorded by the audio meters More detailed information on the true peak level of each channel is displayed within the audio session 7 Max Value The maximum Loudness value recorded wit...

Страница 17: ...sfer of the file 5 The Loudness Session data will be automatically stored in a folder using the Instrument name InstrumentName LoudnessHist Within this folder a file will be created of the Loudness History using the Instrument Name Date and Time of the instrument as the file name 6 Once the data is saved to the memory stick please remember to Unmount the device which can be done under the Main but...

Страница 18: ...ed between the PC and instrument a Tektronix Remote Interface web page should be displayed 10 Select View Loudness History text html from this web page Figure 8 Depending on your network connection speed and the length of the Loudness History this download process can take several minutes 11 The file can be saved to your PC using the web browser application Figure 8 Tektronix Remote Interface Once...

Страница 19: ...or General Knob to select the Loudness Settings menu Figure 9 3 Press SEL to enter sub menus and scroll up and down to select the desired function The following information is provided on each function available within the Loudness Settings menu This information is also available by pressing Help button on the instrument or browsing the On line documentation when connected to the instrument via a ...

Страница 20: ... 85 This preset loads a configuration in compliance with ATSC A 85 and configures the following settings Loudness Filter Measure LKFS BS 1770 Momentary Integration Time IEC 62672 125ms Average Integration Time 10 seconds Channel Weighting ITU R BS 1770 Custom Mask Press SEL Target Loudness 24LKFS Target Loudness High 2LU Target Loudness Low 2LU Gating Type Disabled Safety Gating Level 99LKFS Relat...

Страница 21: ... in compliance with IEC 61672 and configures the following settings Loudness Filter Measure Leq A Momentary Integration Time IEC 62672 125ms Average Integration Time 10 seconds Channel Weighting Custom Custom Mask Press SEL Target Loudness 24dBFS Target Loudness High 2dB Target Loudness Low 2dB Gating Type Disabled Safety Gating Level 99dBFS Relative Gate level 0dB Loud A D Conversion 20dBFS ARIB ...

Страница 22: ...isplaying the loudness ballistic The Program or Channel Pair audio loudness bar displays Loudness measurements and readings in the Audio Display Audio Session Display Audio Loudness Display and Dolby Status Displays The Audio Loudness Session display The currently selected filter is shown within the Audio Display above the Loudness Bar or within the Surround Sound display Momentary Integ Time Aver...

Страница 23: ...when the signal level decreases Loudness F has a 0 125 second time constant IEC62672S This time weighted function applies to the audio level meters and affects how slowly the bars rise when the signal level increases and fall when the signal level decreases Loudness S has a 1 second time constant EBU R 128M This measurement applies the M Momentary ballistic per EBU R 128 with a 400ms integration t...

Страница 24: ...am Loud The loudness level threshold for all channels used by the Loudness measurement above which the loudness value triggers an alarm condition This alarm will trigger an in bar audio message of Too Loud Program Quiet The loudness level threshold for all channels used by the Loudness measurement below which the loudness value triggers an alarm condition This alarm will trigger an in bar audio me...

Страница 25: ...ed within the loudness measurement The default limits is 99LKFS and is user selectable This value is used when Gating type is set to Safety Only Relative Gate Level The loudness level set for Relative Gating measurements When the difference between the current measurement and the overall measurement exceed the relative threshold Then these loudness measurement values will not be included in the ov...

Страница 26: ...0 References WFM WVR8000 Series Waveform Monitors WFM WVR7200 Series Waveform Monitors WFM6000 7000 Series Waveform Monitors WVR6000 7000 Series Waveform Monitors WFM WVR5200 Series Waveform Monitors Data Sheets Fact Sheets and additional product materials can be found www tektronix com video_test signal_monitors html ...

Страница 27: ...How to Guide Audio Loudness Monitoring 27 www tektronix com wfm8000 www tektronix com wfm7100 www tektronix com wfm5200 ...

Страница 28: ...a 65 6356 3900 Austria 00800 2255 4835 Balkans Israel and other ISE Countries 41 52 675 3777 Belgium 00800 2255 4835 Brazil 55 11 3759 7600 Canada 1 800 833 9200 Central East Europe Ukraine and the Baltics 41 52 675 3777 Central Europe Greece 41 52 675 3777 Denmark 45 80 88 1401 Finland 41 52 675 3777 France 00800 2255 4835 Germany 00800 2255 4835 Hong Kong 400 820 5835 India 000 800 650 1835 Ital...

Отзывы: