SECTION 6
Introduction
A complete procedure for checking the operational
standards and calibration of the Type 109 Pulse Generator
is provided in this section of the manual.
Apparent troubles in the instrument are occasionally the
result of an improper adjustment. Consequently, calibration
checks should be an integral part of any troubleshooting
procedure. Abnormal indications occurring during calibra
tion checks will often aid in isolating troubles to a definite
circuit or stage.
In the instructions that follow, the steps are arranged in
proper sequence for a complete check and calibration of
the instrument. Each numbered step contains the informa
tion required to make one check or adjustment or a series of
related checks or adjustments. The steps are arranged to
avoid unnecessary repetition of checks or adjustments.
Test equipment used in a particular step should be left
connected at the end of that step unless instructions state
otherwise. Sim ilarly, controls not mentioned are assumed to
be in positions they were in at the conclusion of the pre
ceding step.
Equipment Required
The following equipment is required to perform a com
plete calibration of the Type 109.
(1) Conventional test oscilloscope such as a Tektronix Type
540-Series Oscilloscope with a Type L Plug-In Unit, a
Type P6000 10X Attenuator Probe and a Type P6027
IX Attenuator Probe. Sustitute specifications: Bandpass
30 me with the 10X probe; deflection factor, at least
50 mv per cm and at least 5 mv per cm at reduced band
pass.
(2) Sampling test oscilloscope such as a Tektronix Type 561
Oscilloscope with a Type 3S76 and a Type 3T77 Plug-In
Units. Substitute specifications: Risetime, 0.4 nsec or less;
any deflection factor up to 20 volts per division. (The
Tektronix Type 519 (conventional) Oscilloscope can
also be used as an excellent substitute.)
(3) One 50-ohm 20-nsec cable, 160", RG-8A/U, with GR
874 connectors on each end, Part No. 017-504.
(4) One 50-ohm 5 nsec cable, 4 0 ", RG-8A/U, with GR
874 connectors on each end, Part No. 017-502.
(5) One 50-ohm terminated cable assembled as shown in
Fig. 6-1. Requires the following parts:
CALIBRATION
F ig . 6 - 1 . 5 0-o h m term in a te d p ig ta il circuit.
1— 8 " RG-58A/U cable with a GR 874 connector on one
end, the other end is pigtailed, Part No. 017-503.
2— 100-ohm V2-watt composition resistor.
F ig . 6 - 2 . M e rc u ry sw itch teste r circuit co nnected to the T yp e 1 09.
(6) One mercury switch teste constructed similar to the one
shown in Fig. 6-2. Use the following parts:
2— 1 V2-volt batteries
1— 22-k y2-watt composition resistor
1— 39-k y2-watt composition resistor
1— 10-k y2-watt composition resistor
3— Banana plugs
1— Alligator clip
1— 3 " wire lead
3— 12"wire leads
@
6-1
Содержание TYPE 109
Страница 8: ...NOTES ...
Страница 12: ...NOTES ...
Страница 22: ...NOTES ...
Страница 28: ...NOTES ...
Страница 34: ...NOTES ...
Страница 44: ...Parts List Type 109 RIGHT SIDE 7 4 ...
Страница 46: ...Parts List Type 109 LEFT SIDE 7 6 ...
Страница 48: ...Parts List Type 109 CABINET 7 8 ...
Страница 50: ...1 3 ...
Страница 54: ... 3 0 3 3 3 3 3 3 3 3 3 3 3 ...
Страница 55: ...TYPE 109 PARTS LIST CORRECTION CHANGE TO SW750 260 0282 02 Mercury Switch Checked M9UL1 L65 ...
Страница 56: ...f 1 Li _r S s s n s i i j n s n j i ...