background image

4

Equipment may not meet the immunity requirements of this standard when test

leads and/or test probes are connected due to coupling of electromagnetic

interference onto those leads/probes. To minimize the in

fl

uence of

electromagnetic interference, minimize the loop area between the unshielded

portions of signal and associated return leads, and keep leads as far away as

possible from electromagnetic disturbance sources. Twisting unshielded test

leads together is an effective way to reduce loop area. For probes, keep the

ground return lead as short as possible and close to the probe body. Some

probes have accessory probe tip adapters to accomplish this most effectively.

In all cases, observe all safety instructions for the probes or leads used.

5

Test port noise

fl

oor can degrade to -70 dBm when the instrument is subjected

to electromagnetic interference per the IEC 61000-4-3 test. The degradation

will be at the speci

fi

c frequencies of the interfering signals.

Australia / New Zealand Declaration of Conformity – EMC

Complies with the EMC provision of the Radiocommunications
Act per the following standard, in accordance with ACMA:

EN 61326-1 and EN 61326-2-1 Radiated and Conducted

Emissions, Class A.

Equipment type

Test and measuring equipment.

Pollution degree description

A measure of the contaminants that could occur in the
environment around and within a product. Typically the
internal environment inside a product is considered to be the
same as the external. Products should be used only in the
environment for which they are rated.

Pollution Degree 2. Normally only dry, nonconductive

pollution occurs. Occasionally a temporary conductivity
that is caused by condensation must be expected. This
location is a typical of

fi

ce/home environment. Temporary

condensation occurs only when the product is out of
service.

Pollution degree rating

Pollution Degree 2 (as de

fi

ned in IEC 61010-1). Rated for

indoor, dry location use only.

Mains overvoltage category rating (AC adapter)

Overvoltage Category II (as de

fi

ned in IEC 61010-1).

Power source speci

fi

cations

Characteristic

Description

AC input voltage

100 V AC — 240 V AC
47 Hz — 63 Hz

DC power output

4.75 V — 5.25 V

VNA power consumption

<16 W

Environmental considerations

Environmental speci

fi

cations

Characteristic

Description

Temperature Range

Operating

5 °C to +50 °C

Non-operating

–40 °C to +71 °C

Humidity

Temperature
Range

Relative
Humidity

+10 °C to +30 °C

5% to 80% RH

+30 °C to +40 °C

5% to 75% RH

Operating
(non-condensing)

+40 °C to +50 °C

5% to 45% RH

Altitude

Operating

3000 meters

Non-operating

4600 meters

Product end-of-life handling

Observe the following guidelines when recycling an instrument
or component:

Equipment recycling.

Production of this equipment required

the extraction and use of natural resources. The equipment may
contain substances that could be harmful to the environment or
human health if improperly handled at the product’s end of life.
To avoid release of such substances into the environment and
to reduce the use of natural resources, we encourage you to
recycle this product in an appropriate system that will ensure
that most of the materials are reused or recycled appropriately.

This symbol indicates that this product complies
with the applicable European Union requirements
according to Directives 2012/19/EU and
2006/66/EC on waste electrical and electronic
equipment (WEEE) and batteries. For information
about recycling options, check the Tektronix Web
site (www.tek.com/productrecycling).

Front panel

No.

Component

Description

1

RF Port 1

N-type 50

female test port for

device-under-test (DUT) connection.
Use either RF port for stimulus source or
response receiver.

2

LO A

SMA 50

female port 10 dBm, 0 V DC

(reserved for future use)

3

LED
Indicator

Indicates whether the VNA is powered
and disconnected (red) or connected
(green).

4

LO B

SMA 50

female port 10 dBm, 0 V DC

(reserved for future use)

5

RF Port 2

N-type 50

female test port for

device-under-test (DUT) connection.
Use either RF port for stimulus source or
response receiver.

Rear panel

No.

Component

Description

1

Bias 2

Provides bias input for RF port 2

2

Aux Sync

50

female BNC connector (reserved

for future use)

3

Trigger In

50

female BNC connector to connect

an external device to provide a trigger
input

4

USB 2.0

Connect USB 2.0 cable

5

4.75V—5.25V
DC

DC input for the instrument

6

Ref In

50

female BNC connector to provide

optional 10 MHz reference input

7

Ref Out

50

female BNC connector to provide

10 MHz reference output

8

Bias 1

Provides bias input for RF port 1

Copyright © Tektronix, Inc. All rights reserved.

www.tek.com

Содержание TTR500 series

Страница 1: ...before you use it Look for cracks or missing pieces Use only specified replacement parts Do not operate in wet damp conditions Be aware that condensation may occur if a unit is moved from a cold to a...

Страница 2: ...nsumption 16 W Environmental considerations Environmental specifications Characteristic Description Temperature Range Operating 5 C to 50 C Non operating 40 C to 71 C Humidity Temperature Range Relati...

Страница 3: ...MC EC EMC 2014 30 EU Of cial Journal of the European Communities EN 61326 1 EN 61326 2 1 EMC 1 2 3 4 CISPR 11 1 A IEC 61000 4 2 IEC 61000 4 3 RF 5 IEC 61000 4 4 IEC 61000 4 5 IEC 61000 4 6 RF IEC 6100...

Страница 4: ...C 5 80 30 C 40 C 5 75 40 C 50 C 5 45 3 000m 4 600m WEEE 2012 19 EC 2006 66 EC EU Web www tek com productrecycling 1 RF 1 DUT N 50 RF 2 LO A SMA 50 10dBm 0V DC 3 LED VNA 4 LO B SMA 50 10dBm 0V DC 5 RF...

Страница 5: ...C EC EMC Directive 2014 30 EU EN 61326 1 EN 61326 2 1 EMC 1 2 3 4 CISPR 11 1 A IEC 61000 4 2 IEC 61000 4 3 RF 5 IEC 61000 4 4 IEC 61000 4 5 IEC 61000 4 6 RF IEC 61000 4 11 EN 61000 3 2 AC EN 61000 3 3...

Страница 6: ...C 5 75 RH 40 C 50 C 5 45 RH 3 000m 4 600m WEEE Directive 2012 19 EU 2006 66 EC www tek com productrecycling 1 RF 1 DUT N 50 RF 2 LO A SMA 50 10dBm 0V DC 3 LED VNA 4 LO B SMA 50 10dBm 0V DC 5 RF 2 DUT...

Страница 7: ...xx Tektronix Tektronix Inc 14150 SW Karl Braun Drive P O Box 500 Beaverton OR 97077 USA 1 800 833 9200 www Tek com II III IV Tektronix c TTR500 x 4 071 3486 00...

Страница 8: ...tek com 1 2 3 4 5 70 IEC 61000 4 3 ACMA EN 61326 1 EN 61326 2 1 A 2 2 IEC 61010 1 II IEC 61010 1 100 240 47 63 4 75 5 25 16 5 50 C 40 71 C 10 30 C 5 80 30 40 C 5 75 40 50 C 5 45 3000 4600 2012 19 EU...

Страница 9: ...EMC EMC EC EMC Directive 2014 30 EU EN 61326 1 EN 61326 2 1 EMC 1 2 3 4 CISPR 11 1 A IEC 61000 4 2 IEC 61000 4 3 5 IEC 61000 4 4 IEC 61000 4 5 IEC 61000 4 6 IEC 61000 4 11 EN 61000 3 2 EN 61000 3 3 Te...

Страница 10: ...C 71 C 10 C 30 C 5 80 RH 30 C 40 C 5 75 RH 40 C 50 C 5 45 RH 3000 4600 WEEE 2012 19 EU 2006 66 EC Tektronix www tek com productrecycling 1 1 N 50 DUT 2 LO A SMA 50 10 dBm 0 V DC 3 LED VNA 4 LO B SMA...

Страница 11: ...EMC EC EMC 2014 30 EU EN 61326 1 EN 61326 2 1 EMC 1 2 3 4 CISPR 11 1 A IEC 61000 4 2 IEC 61000 4 3 RF 5 IEC 61000 4 4 IEC 61000 4 5 IEC 61000 4 6 RF IEC 61000 4 11 EN 61000 3 2 EN 61000 3 3 Mfr Tektr...

Страница 12: ...6 66 EC Tektronix www tek com productrecycling 1 RF Port 1 RF 1 DUT N 50 RF 2 LO A SMA 50 10 dBm 0 V DC 3 LED Indicator LED VNA 4 LO B SMA 50 10 dBm 0 V DC 5 RF Port 2 RF 2 DUT N 50 RF 1 Bias 2 2 RF 2...

Страница 13: ...v Kh ng ch m v o c c k t n i v b ph n kh ng c b o v khi ang c i n Kh ng v n h nh khi nghi ng c h h ng N u b n nghi ng r ng s n ph m n y c h h ng h y a s n ph m cho nh n vi n d ch v c tr nh ki m tra V...

Страница 14: ...ph m c nh gi M c nhi m 2 Th ng th ng nhi m ch x y ra m i tr ng kh kh ng d n i n Th nh tho ng t nh d n i n t m th i c g y ra b i s ng ng t ph i c d b o a i m n y g m m i tr ng v n ph ng gia nh i n h n...

Отзывы: