Installation
GPIB interface
The GPIB connector supports the IEEE-488.1 (1978) interface standard. It also
supports the required common commands of the IEEE-488.2 (1987) standard.
NOTE.
Changes to the GPIB con
fi
guration do not take effect until the interface is
reset or the instrument power is cycled.
To set up the GPIB interface:
1.
Connect a GPIB cable to the rear panel GPIB port.
2.
Press the
Utility
button from the main menu.
3.
Select
I/O Interface
.
4.
Select
GPIB
.
5.
Check that GPIB is set to
ON
. If it is not, highlight
ON
in the GPIB menu
to turn it on.
6.
Press the
Address
button and use the number keys to assign a unique address
to the instrument.
NOTE.
Each device connected to the GPIB bus must have a unique GPIB address.
The GPIB address must be from 0 to 30.
7.
Press the
Enter
button.
8.
Press the
Return
button three times to return to the main menu.
RS-232 interface
The RS-232 interface connector is a standard 9 pin, type D, female connector
con
fi
gured as a DCE (transmit on pin 2, receive on pin 3). In order to communicate
properly over RS-232, the instrument and the host computer both must be set up
to use the same con
fi
guration. The following baud rates are supported: 115200
(default), 57600, 38400, 19200, 9600, and 4800. The rest of the communication
parameters are
fi
xed at 8 data bits, 1 stop bit, no parity, and RTS/CTS hardware
fl
ow control.
To set up the RS-232 interface:
1.
Connect a cable to the rear panel RS-232 port. Use a cable best suited for the
desired baud rate (modulation rate).
2.
Press the
Utility
button from the main menu.
3.
Select
I/O Interface
.
4.
Select
RS232
.
TSG4100A Series RF Signal Generators User Manual
5
Содержание TSG4102A
Страница 1: ...xx TSG4100A Series RF Signal Generators ZZZ User Manual P071315002 071 3150 02...
Страница 2: ......
Страница 16: ...Preface xii TSG4100A Series RF Signal Generators User Manual...
Страница 87: ...Reference Reference Phase noise and offset diagrams TSG4100A Series RF Signal Generators User Manual 71...