Performance veri
fi
cation
Performance veri
fi
cation
The procedures that follow verify the warranted speci
fi
cations of the probe. The recommended calibration interval is one
year. Perform the veri
fi
cation procedures in the order listed.
Equipment required
The following equipment is required for the performance veri
fi
cation procedures.
Table 8: Test equipment
Description and quantity
Performance requirement
Recommended example
1
Oscilloscope
TekVPI Interface
Tektronix 6 Series MSO
8 Hz bandwidth option
DC calibration source
Keithley 2400 SMU
Digital multimeter (DMM)
Resistance, 0.1% accuracy
Keithley 2700 DMM
Network Analyzer
Tektronix VNA TTR506A
067-1701-XX with Calibration kit
BN533828
TekVPI Calibration Veri
fi
cation adapter
TekVPI Interface
SMA to BNC adapter
SMA male to BNC female
015-0554-XX
SMA to BNC adapter
SMA female to BNC male
015-0572-XX
SMA to SMA adapter
SMA male to SMA male
015-0551-XX
BNC-to-dual banana adapter (2)
103-0090-XX
BNC cable (2)
50
Ω
, 0.76 m (30 in) length
012-0117-XX
Feed-thru termination
50
Ω
, 1 GHz, ±0.5
Ω
011-0049-XX
SMA cable for network analyzer
N to SMA-M 5 foot cable
012-1774-XX
SMA adapter for network analyzer
Type-N male to Type-SMA female
013-0406-XX
SMA torque wrench
5/16-in, 7 in-lb.
SMA adapter wrench
7/32-in
1
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers.
26
TPR1000 and TPR4000 User Manual