Introduction
Functional Check
After you have installed the battery or connected external power, perform this
quick functional check to verify that your instrument is operating correctly. See
the
THS3000 Series Oscilloscopes Installation and Safety Instructions
for battery
installation and power information.
Perform this adjustment to match your probe to the input channel. This should be
done whenever you attach a passive probe for the
fi
rst time to any input channel.
The calibration consists of a high frequency adjustment and a DC calibration for
10:1 probes and 100:1 probes. You can also read more about probe calibration.
(See page 103,
Probe Compensation and Compatible Maximum Voltages
1.
Press the power button on the front panel to turn on the instrument.
2.
After a few seconds, you should see a colored trace on the display screen. The
color of the trace corresponds to the color of the channel input button on
the front panel.
3.
Connect a voltage probe to the Ch 1 input BNC on the instrument top panel.
The color at the base of each BNC matches the color of the associated channel
button and the color of the probe.
4.
Attach the probe tip and reference lead to the probe calibration connector
located above the USB port on the left side of the instrument.
NOTE.
Connect the probe tip to the smallest metal connector (on the top) and the
reference lead to the larger metal connector (on the bottom).
5.
Press the
Ch 1
button to view the menu.
6.
Select
PROBE 1
and then use the arrow keys and the
Enter
button to choose
the appropriate attenuation from the menu.
7.
Return to the
PROBE 1
menu and select
PROBE CAL
.
8.
Select
Yes
.
9.
You will see the rising edge of a square wave on the display (approximately
500 Hz). This is the signal used to calibrate the probe. (See Figure 2.) (See
page 103,
Probe Compensation and Compatible Maximum Voltages
THS3000 Series Oscilloscopes User Manual
11
Содержание THS3014
Страница 1: ...xx THS3000 Series Oscilloscopes ZZZ User Manual P077060702 077 0607 02...
Страница 2: ......
Страница 3: ...THS3000 Series Oscilloscopes ZZZ User Manual xx www tektronix com 077 0607 02...
Страница 6: ......
Страница 14: ...General safety summary viii THS3000 Series Oscilloscopes User Manual...
Страница 32: ...Introduction 16 THS3000 Series Oscilloscopes User Manual...
Страница 75: ...Analyze Waveforms Figure 30 Risetime measurement using cursors THS3000 Series Oscilloscopes User Manual 59...
Страница 96: ...Troubleshooting 80 THS3000 Series Oscilloscopes User Manual...
Страница 112: ...Appendix A Specifications 96 THS3000 Series Oscilloscopes User Manual...
Страница 118: ...Appendix B Connect the Instrument to a Computer 102 THS3000 Series Oscilloscopes User Manual...
Страница 122: ...Appendix C Probe Compensation and Compatible Maximum Voltages 106 THS3000 Series Oscilloscopes User Manual...
Страница 126: ...Appendix D Battery Pack 110 THS3000 Series Oscilloscopes User Manual...
Страница 128: ...Appendix E Maintenance and Cleaning 112 THS3000 Series Oscilloscopes User Manual...