Operating considerations
Operating considerations
Table 1: TDP0500 & TDP1000
Characteristic
Description
Input Voltage
Differential:
±
4.25 V (DC + peak AC), 3 V RMS
±
42 V (DC + peak AC), 30 V RMS
Common Mode:
±
35 V (DC + peak AC), 25 V RMS (Both ranges; input referenced to ground)
Temperature
Operating: 0 to +40 °C (+32 °F to +104 °F)
Nonoperating: -55 °C to +75 °C (-67 °F to +167 °F)
Humidity
Operating: +30 °C to +40 °C (+86 °F to +104 °F) 0-90% RH
Nonoperating: +30 °C to +60 °C (+86 °F to +140 °F) 0-90% RH
Altitude
Operating: Up to 3000 meters (10,000 feet)
Nonoperating: Up to 12192 meters (40,000 feet)
Pollution Degree
2, Indoor use only