Performance Veri
fi
cation
Equipment Required
The equipment required for the performance veri
fi
cation procedures are listed below.
Table 6: Test equipment
Description and quantity
Performance requirement
Recommended example
1
Oscilloscope
TekVPI interface, 500 MHz or greater bandwidth
Tektronix DPO4000
High amplitude pulse generator
Rise time <500 ps, pulse width >100 ns, amplitude
>10 Vpp into 50
Ω
Picosecond Labs 2600
Calibrator
DCV: 0.2% accuracy, 0 to ±1.5 V, square wave output
ACA: 0.25% accuracy, 0 to ±6 A, square wave output
Fluke 9100
Digital Multimeter (DMM)
0.02% accuracy
Tektronix DMM4040/4050
Adapter
TekVPI calibration/veri
fi
cation adapter
067-1701-xx
Adapter
TekVPI-to-BNC interface adapter
TPA-BNC
DC current loop
5 turns 18 AWG coated wire on 3 inch form
See instructions that follow
HF current loop
50
Ω
±0.5%, BNC male
067-2396-xx
Adapter
BNC-to-dual banana plug
103-0090-xx
Adapter
SMA male-to-BNC-female
015-1018-xx
BNC cable
50
Ω
, 0.76 m (30 in) length
012-0117-xx
1
Nine-digit part numbers (xxx-xxxx-xx) are Tektronix part numbers.
48
TCP202A Current Probe Instruction Manual
Содержание TCP202A
Страница 1: ...x TCP202A 50 MHz 15 A AC DC Current Probe ZZZ Instruction Manual P071300300 071 3003 00...
Страница 2: ......
Страница 6: ......
Страница 10: ...Table of Contents iv TCP202A Current Probe Instruction Manual...
Страница 62: ...Specifications Figure 2 Typical input impedance versus frequency 42 TCP202A Current Probe Instruction Manual...
Страница 63: ...Specifications Figure 3 Maximum peak pulse versus pulse width TCP202A Current Probe Instruction Manual 43...
Страница 65: ...Specifications Figure 4 Probe dimensions TCP202A Current Probe Instruction Manual 45...