Connecting the Probe
Degaussing the Probe
1.
To degauss the probe,
fi
rst verify that the
probe jaw is fully closed.
2.
Press the
DEGAUSS
button on the side
of the probe.
Quick Tip
To maintain measurement accuracy, degauss
your probe in each of these cases:
After you turn on the measurement system
and allow a 20-minute warm-up period
Before you connect the probe to a
conductor
Whenever you subject the probe to a
strong external magnetic
fi
eld
TCP202A Current Probe Instruction Manual
5