Performance Veri
fi
cation
24.
Use the measured waveform amplitude (e)
from step 23 and the oscilloscope DC gain
accuracy value (b) from step 7 to calculate
the probe DC gain accuracy (f).
25.
Record the calculated probe DC gain
accuracy value (f) in the test record.
30 A Range:
Rise Time and Bandwidth
This procedure veri
fi
es that the probe meets the rise time speci
fi
cation on both of the current ranges. The bandwidth of the
probe is then calculated using the measured probe rise time.
TCP0030 Current Probe Instruction Manual
59