Trigger setup
Setting the trigger mode
Use this procedure to set the oscilloscope trigger
Mode
. The trigger mode sets how the oscilloscope behaves in the absence
or presence of a trigger. Trigger mode also enables the trigger holdoff function. (See page 28,
Trigger mode does not set the trigger conditions; those are set with the
Trigger Type
side-menu buttons.
1.
Push the
Trigger Menu
front-panel button.
2.
Push the
Mode
side-menu button.
3.
Use the
Multipurpose
knob to select and click a trigger mode or holdoff mode:
Click
Auto (Untriggered Roll)
to set the oscilloscope to acquire and display a waveform even if a trigger does not
occur. If there is no valid triggering event, the waveform will appear to roll across the screen.(See page 28,
(Untriggered Roll) trigger mode
Click
Normal
to set the oscilloscope to acquire and display a waveform only when a valid trigger event occurs. (See
Click
Holdoff
to set a time period after a trigger event, during which the oscilloscope ignores new trigger events. Once
the holdoff period is over, the oscilloscope starts looking for the trigger condition. The minimum holdoff period is
20 ns. (See page 29,