Before installation
Operating requirements
Power supply requirements
The power supply requirements for your instrument and controller are listed in the following table.
WARNING.
To reduce the risk of
fi
re and shock, ensure that the mains supply voltage
fl
uctuations do not exceed 10% of the
operating voltage range.
RSA7100
CRTL7100
Source voltage and frequency
100 V AC to 240 V AC, 50/60 Hz
100 V AC to 240 V AC, 50/60 Hz
Power consumption
400 W maximum. Maximum line
current is 4.5 A at 90 V line.
300 W, typical.
500 W maximum. Maximum line current is
5.5 A at 90 V line.
400 W, typical.
Over current protection.
A 15 A circuit breaker located on the instrument rear panel will trip if an over current condition
is encountered. This will help prevent damage to the instrument. You can reset the circuit breaker by
fi
rst disconnecting the
power cable, pressing the circuit breaker to reset it, and then reconnecting the power cable.
Environmental requirements
The environmental requirements for the instrument and controller are listed in the following table. For instrument accuracy,
ensure that the instrument has warmed up for 20 minutes and meets the environmental requirements listed here.
Requirement
RSA7100
CTRL7100
Temperature
Operating
0 °C to +40 °C (+32 °F to +104 °F)
+10 °C to +35 °C (+50 °F to +95 °F)
Non-operating
–20 °C to +60 °C (–4 °F to +140 °F)
–20 °C to +60 °C (–4 °F to +140 °F)
Relative humidity, non-condensing
Operating
10% to 90% up to 40 °C
40% to 70% up to 35 °C
Altitude
Operating
Up to 2,000 m (6,561 feet)
Up to 3,000 m (9,842 feet)
Non-operating
Up to 12,000 m (39,370 feet)
Up to 12,000 m (39,370 feet)
Cooling
Clearance requirements for the RSA7100.
Observe these clearance requirements for the RSA7100 when placing it
on a cart, bench, or rack.
Top/bottom: 44.45 mm (1.75 in)
Sides: 44.45 mm (1.75 in)
Rear: 76.2 mm (3.0 in)
Ensure that no outside fans are blowing into the RSA7100 rear fans.
4
RSA7100 Series Quick Start User Manual
Содержание RSA7100 Series
Страница 1: ...x RSA7100 Series Real Time Spectrum Analyzer ZZZ Quick Start User Manual 071 3504 10 ...
Страница 2: ......
Страница 6: ......
Страница 20: ...Preface xiv RSA7100 Series Quick Start User Manual ...
Страница 102: ...Operation 82 RSA7100 Series Quick Start User Manual ...