![Tektronix RSA6100A Series Скачать руководство пользователя страница 99](http://html1.mh-extra.com/html/tektronix/rsa6100a-series/rsa6100a-series_help-manual_1078169099.webp)
General Signal Viewing > Spectrogram
Time & Freq Scale Tab
Setting
Description
Vertical (Time)
Normal
For most Spectrogram applications. Primary time scale control is Time/div. Time scale
can be zoomed in or out.
Spectrum Monitor
For long-term signal monitoring applications. In spectrum monitor mode, each line in the
spectrogram represents the period of time speci
fi
ed by the Time/update parameter. Time
scale can be zoomed out, but not zoomed in (no overlap).
Time/update
Sets the time, in minutes and seconds, represented by each line of the spectrogram.
Only available in Spectrum Monitor.
Reset Scale
Resets the Time/div and Pos settings to their default values.
Autoscale
Scales the vertical (time) axis to compress all existing trace lines into the visible area
of the graph. Resets the Position value to zero, placing the most recent spectrogram
line at the bottom of the spectrogram display. Only Position is affected by Autoscale
when Spectrum Monitor is selected.
Time/div
Sets the time displayed per division.
Visible elapsed time
Displays the length of time visible in the display. This does not represent the total time
available to view.
Position
Adjusts vertical position of the trace within the graph area. Setting represents the offset,
in divisions, between the bottom of the graph and the bottom (most recent) line in the
results trace.
Time at position
Displays the time of the spectrogram line shown at the bottom of the graph. This time is
relative to the Time Zero Reference of the current acquisition.
Horizontal (Frequency)
Scale
Sets the frequency range of the graph without changing the Span value.
Position
Sets the frequency displayed at the center of the graph. Changing this value does not
change the Freq setting.
Autoscale
Sets the frequency scale to the Spectrogram Span value.
xxx
Spectrum Monitor
Spectrum Monitor performs long term monitoring. The monitor mode compresses time into each line
of the spectrogram, which enables you to monitor long periods of time (from 1 second per line up to
600 minutes per line). With extended memory (Option 02) installed in the instrument, you can capture up
to 125,000 lines. With standard memory installed in the instrument, the maximum number of lines you
can capture is 31,250.
During each line’s collection period, spectrum transforms are computed for each acquisition taken by the
instrument. As each transform completes, it is incorporated into the current spectrogram line. How each
RSA6100A Series Real-Time Spectrum Analyzers Help
87
Содержание RSA6100A Series
Страница 1: ...xx RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document P077016903 077 0169 03...
Страница 2: ......
Страница 3: ...RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document www tektronix com 077 0169 03...
Страница 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 285: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 273...
Страница 286: ...GP Digital Modulation Symbol Maps Symbol Maps 274 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 287: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 275...
Страница 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...
Страница 388: ...Reference Troubleshooting Displaying the Windows Event Viewer 376 RSA6100A Series Real Time Spectrum Analyzers Help...