RF Measurements > Common Controls for Settling Time
Displays
Mask Tab for Settling Time Displays
Mask Tab for Settling Time Displays
The Mask tab is used to specify the parameters used for Mask testing.
Enable Mask Test
Enables or disables mask testing.
Start 1, 2, 3
Sets the starting point for the time zone of interest. If you don’t want to use all three zones, set the Start
Time for the unneeded zones to a time after the end of the Measurement Length.
Stop
Sets the end point for the
fi
nal time zone of interest.
Limit (+/-)
Sets the mask violation limits for each time zone.
Mask Time Reference
When mask testing is enabled, the Mask time reference selection speci
fi
es the starting point of the mask.
You can choose to start the mask test at the trigger point or at the point on the trace where the signal began
its transition from its previous settled frequency or phase level to the level you intend to measure.
This capability is useful because some applications do not have a trigger signal available. For those
situations where a trigger signal is available, the speci
fi
cation for the device-under-test may require
that measurements be referenced to that signal.
Trace Tab for Settling Time Displays
The Traces Tab allows you to set the display characteristics of displayed traces.
RSA6100A Series Real-Time Spectrum Analyzers Help
193
Содержание RSA6100A Series
Страница 1: ...xx RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document P077016903 077 0169 03...
Страница 2: ......
Страница 3: ...RSA6100A Series Real Time Spectrum Analyzers ZZZ Printed Help Document www tektronix com 077 0169 03...
Страница 12: ...Table of Contents viii RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 28: ...Orientation Elements of the Display 16 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 285: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 273...
Страница 286: ...GP Digital Modulation Symbol Maps Symbol Maps 274 RSA6100A Series Real Time Spectrum Analyzers Help...
Страница 287: ...GP Digital Modulation Symbol Maps Symbol Maps RSA6100A Series Real Time Spectrum Analyzers Help 275...
Страница 307: ...Search Limits Testing Define Tab Search RSA6100A Series Real Time Spectrum Analyzers Help 295...
Страница 388: ...Reference Troubleshooting Displaying the Windows Event Viewer 376 RSA6100A Series Real Time Spectrum Analyzers Help...