Commands listed in alphabetical order
this command speci
fi
es the lower limit of the range. (Use the command
TRIGger:{A|B}:BUS:B<x>:ETHERnet:DATa:HIVALue
to set the upper limit of
the range.) The default is all X’s (don’t care). The bus number is speci
fi
ed by <x>.
Conditions
Requires the SR-ENET Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:ETHERnet:DATa:VALue <QString>
TRIGger:{A|B}:BUS:B<x>:ETHERnet:DATa:VALue?
Related Commands
TRIGger:{A|B}:BUS:B<x>:ETHERnet:CONDition
TRIGger:{A|B}:BUS:B<x>:ETHERnet:DATa:OFFSet
TRIGger:{A|B}:BUS:B<x>:ETHERnet:DATa:SIZe
TRIGger:{A|B}:BUS:B<x>:ETHERnet:DATa:HIVALue
Arguments
QString is a quoted string where the allowable characters are 0, 1, and X.
The allowable number of characters depends on the setting for size (using
TRIGger:A:BUS:B<x>:ETHERnet:DATa:SIZe). The bits speci
fi
ed in the
quoted string replace the least signi
fi
cant bits, leaving any unspeci
fi
ed upper
bits unchanged.
Examples
TRIGger:A:BUS:B1:ETHERnet:DATa:VALue "00001000"
sets the binary
data to trigger on to 00001000, assuming the quali
fi
er is set to
LESSthan,
MOREthan, EQual, UNEQual, LESSEQual
or
MOREEQual,
and
DATa:SIZe
is set to 1 byte.
TRIGger:A:BUS:B1:ETHERnet:DATa:VALue "00001000"
sets the lower
limit of the range to 00001000, assuming the quali
fi
er is set to
INrange
or
OUTrange
, and
DATa:SIZe
is set to 1 byte.
TRIGger:{A|B}:BUS:B<x>:ETHERnet:IPHeader:DESTinationaddr:VALue
When the Ethernet trigger condition is set to
IPHeader
, this command speci
fi
es
the value of the 32–bit destination address that is to be used in the trigger (along
with the source address and protocol value). The IP destination address is a
standard IP address such as 192.168.0.1. The default is all X’s (don’t care). The
bus number is speci
fi
ed by <x>.
Conditions
Requires the SR-ENET Triggering and Analysis application.
MSO54, MSO56, MSO58, MSO58LP Programmer
2-783
Содержание MSO54
Страница 1: ...xx 5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual P077130502 077 1305 02 ...
Страница 2: ......
Страница 3: ...5 Series MSO MSO54 MSO56 MSO58 MSO58LP ZZZ Programmer Manual xx www tek com 077 1305 02 ...
Страница 8: ...Preface iv MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 26: ...Command Syntax 2 10 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 936: ...Status and Events 3 20 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 937: ...Appendix A Character Set MSO54 MSO56 MSO58 MSO58LP Programmer A 1 ...
Страница 938: ...Appendix A Character Set A 2 MSO54 MSO56 MSO58 MSO58LP Programmer ...
Страница 996: ...Appendix D Waveform Transfer WFMOutpre and CURVe Query Examples D 18 MSO54 MSO56 MSO58 MSO58LP Programmer ...