Installation
3.
Connect the probe connector to
oscilloscope channel 1 and the probe tip
and reference lead to the
PROBE COMP
terminals on the oscilloscope front panel.
4.
Push
Default Setup
.
5.
Push
Autoset
. The screen should now
display a square wave, approximately
2.5 V at 1 kHz.
If the signal appears but is misshapen,
perform the procedures for compensating
the probe.
If no signal appears, rerun the procedure.
If this does not remedy the situation,
have the instrument serviced by quali
fi
ed
service personnel.
Compensating a TPP0250, TPP0500B or TPP1000 Passive Voltage Probe
The MDO3000 Series oscilloscopes can automatically compensate TPP0250, TPP0500B and TPP1000 probes. This
eliminates the need for manual probe compensation, as is typically performed with other probes.
Each compensation generates values for a speci
fi
c probe and channel combination. If you want to use the probe on another
channel and desire to compensate the new probe-channel pair, you must run a new set of compensation steps for that new
combination.
1.
Connect the oscilloscope power cable. (See
page 9,
2.
Power on the oscilloscope.
MDO3000 Series Oscilloscopes User Manual
11