Trigger Setup
Checking Trigger Settings
To quickly determine the settings of some
key trigger parameters, check the Trigger
readout at the bottom of the display. The
readouts differ for edge and the advanced
triggers.
1.
Trigger source = channel 1.
2.
Trigger slope = rising.
3.
Trigger level = 0.00 V.
Edge trigger readout
Using Sequence Trigger (A (Main) and B (Delayed))
Combine an edge A Event (Main) trigger with the B Event (Delayed) trigger to capture more complex signals. After the A
Event occurs, the trigger system looks for the B Event before triggering and displaying the waveform.
A and B triggers can (and typically do) have separate sources.
NOTE.
You can select sequence triggering when you choose the slope type
Falling
or
Rising
— but not when you pick
the slope type
Both
.
Use the Edge trigger menu to set up the A trigger
fi
rst. Then, to use the B trigger:
1.
Push Trigger
Menu
.
2.
Push
Type
.
3.
Turn
Multipurpose a
to select a trigger type
of
Sequence (B Trigger)
.
This brings up the Sequence (B Trigger)
menu.
4.
Push
B Trigger After A
.
Type
Sequence
(B Trigger)
Source
1
Coupling
DC
Slope
Level
0.00 V
B Trigger
After A
Time
Mode
Auto
& Holdoff
100
MDO3000 Series Oscilloscopes User Manual
Содержание MDO3012
Страница 1: ...x MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual P077096802 077 0968 02...
Страница 2: ......
Страница 3: ...MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tek com 077 0968 02...
Страница 10: ...Table of Contents iv MDO3000 Series Oscilloscopes User Manual...
Страница 236: ...Appendix D OpenSSL License 214 MDO3000 Series Oscilloscopes User Manual...