Appendix A: Common commands
2470 High Voltage SourceMeter Instrument Reference Manual
A-6
2470-901-01 Rev. A /
May
2019
*LANG
This command determines which command set is used by the instrument.
Type
Affected by
Where saved
Default value
Command and query
Not applicable
Nonvolatile memory
SCPI
Usage
*LANG <commandSet>
*LANG?
<commandSet>
The command set to be used:
TSP
SCPI
Details
The remote command sets that are available include:
•
SCPI: An instrument-specific language built on the SCPI standard.
•
TSP: A scripting programming language that contains instrument-specific control commands that
can be executed from a stand-alone instrument. You can use TSP to send individual commands
or use it to combine commands into scripts.
If you change the command set, reboot the instrument.
You cannot combine the command sets.
Example
*LANG TSP
*LANG?
Set the command set to TSP.
Verify setting by sending the command set query.
Output:
TSP
The TSP command set is in use.
Also see
(on page B-1)
*OPC
This command sets the operation complete (OPC) bit after all pending commands, including overlapped
commands, have been executed.
Type
Affected by
Where saved
Default value
Command and query
Not applicable
Not applicable
Not applicable
Usage
*OPC
*OPC?