2470 High Voltage SourceMeter Instrument Reference Manual
Section 14: TSP command reference
2470-901-01 Rev. A /
May
2019
14-213
Details
The LAN trigger listens for trigger messages on all the supported protocols. However, it uses the
designated protocol for sending outgoing messages.
After you change this setting, you must re-connect the LAN trigger event generator before you can
send outgoing event messages.
When multicast is selected, the trigger IP address is ignored and event messages are sent to the
multicast address 224.0.23.159.
Example
print(trigger.lanout[1].protocol)
Get LAN protocol that is being used for sending trigger
messages for LAN event 1.
Also see
(on page 14-209)
trigger.lanout[N].stimulus
This attribute specifies events that cause this trigger to assert.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script
trigger.EVENT_NONE
Usage
LANevent
= trigger.lanout[
N
].stimulus
trigger.lanout[
N
].stimulus =
LANevent
LANevent
The LAN event that causes this trigger to assert; see
Details
for values
N
A number specifying the trigger packet over the LAN for which to set or query the
trigger source (1 to 8)
Details
This attribute specifies which event causes a LAN trigger packet to be sent for this trigger. Set the
event to one of the existing trigger events, which are shown in the following table.