Section 14: TSP command reference
2470 High Voltage SourceMeter Instrument Reference Manual
14-204
2470-901-01 Rev. A /
May
2019
trigger.digout[N].release()
This function releases an indefinite length or latched trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.digout[
N
].release()
N
Digital I/O trigger line (1 to 6)
Details
Releases a trigger that was asserted with an indefinite pulsewidth time. It also releases a trigger that
was latched in response to receiving a synchronous mode trigger. Only the specified trigger line is
affected.
Example
digio.line[4].mode = digio.MODE_TRIGGER_OUT
trigger.digout[4].release()
Releases digital I/O trigger line 4.
Also see
(on page 14-52)
Digital I/O port configuration
(on page 14-201)
(on page 14-203)
trigger.digout[N].stimulus
This attribute selects the event that causes a trigger to be asserted on the digital output line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Digital I/O trigger
N
reset
Configuration script
trigger.EVENT_NONE
Usage
event
= trigger.digout[
N
].stimulus
trigger.digout[
N
].stimulus =
event
event
The event to use as a stimulus; see
Details
N
Digital I/O trigger line (1 to 6)