2470 High Voltage SourceMeter Instrument Reference Manual
Section 12: SCPI command reference
2470-901-01 Rev. A /
May
2019
12-77
:SOURce[1]:<function>[:LEVel][:IMMediate][:AMPLitude]
This command immediately selects a fixed amplitude for the selected source function.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Source configuration list
Save settings
Source configuration list
0
Usage
:SOURce[1]:<function>[:LEVel][:IMMediate][:AMPLitude] <value>
:SOURce[1]:<function>[:LEVel][:IMMediate][:AMPLitude] <DEF|MIN|MAX>
:SOURce[1]:<function>[:LEVel][:IMMediate][:AMPLitude]?
:SOURce[1]:<function>[:LEVel][:IMMediate][:AMPLitude]? <DEF|MIN|MAX>
<function>
The source function to which this setting applies:
Current:
CURRent
Voltage:
VOLTage
<value>
The amplitude:
Current:
−
1.05 A to 1.05 A
Voltage:
−
1100 V to 1100 V
<DEF|MIN|MAX>
The
DEFault
,
MINimum
, or
MAXimum
value
Details
This command sets the output level of the voltage or current source. If the output is on, the new level
is sourced immediately.
The sign of the source level dictates the polarity of the source. Positive values generate positive
voltage or current from the high terminal of the source relative to the low terminal. Negative values
generate negative voltage or current from the high terminal of the source relative to the low terminal.
If a manual source range is selected, the level cannot exceed the specified range. For example, if the
voltage source is on the 2 V range, you cannot set the voltage source level to 3 V. When autorange is
selected, the amplitude can be set to any level supported by the instrument.
Example
SOUR:FUNC VOLT
SOUR:VOLT 1
Set the instrument to source voltage and set
it to source 1 V.
Also see
(on page 12-82)